HS/KS260 0305
ATTENTION!
After a power failure, the
unlocked (freely movable) agitation table will con-
tinue to perform a circular motion until the lock
function can fasten the agitation table in place.
Basic model
The amount of time required for the agitation process can be set
with the aid of the right turn button. A distinction is made for the
agitation time setting between timed mode and continuous ope-
ration. If continuous mode is selected (On LED is lit), the agita-
tor can continue its agitation function for any amount of time
with the previously set speed.
If timed mode is selected, the agitation process can last for a
maximum of 50 min. The set time is displayed by the LED line.
The flashing LED shows the amount of time currently remaining
After all LED’s have gone out in timed mode, the device swit-
ches off into the Off operating mode and the agitation process
comes to an end.
Note:
The curently set time can be modified at any time. If the
time setting is set to the right beyond 50 minutes, the LED bar
goes out and the LED in the On operating state (continuous
mode) starts to light up. If you continue to hold down the turn
button, the device first jumps to the Off operating mode (stand-
still) and then back into timed operation. The same applies in the
opposite order if the turn button is held down beyond a certain
amount of time and moved to the left.
After the power failed, the set time (timed mode or continuous
mode) is erased and the device switches into Offline operating
mode (standstill).
audio signal indicating that time
has expired by pressing the right
turn button.
Note:
If the time expiration signal has been activated, expiration
of the set time is indicated by a sequence of signals (3 times 5
short tons at intervals of about 10 seconds).
Adjusting the lock function
The lock function of the agitation table can be activated or
deactivated with a direct connection to the setting of the
audio time expiration signal.
The following steps are required to activate or deactivate
the function:
a.)
Turn off the device with the
power switch.
b.)
Hold down the right turn
button and turn on the devi-
ce with the power switch.
Let go of the turn button
after about 5 seconds. After
you have run through the
Sound
On/Off
routine, it is
possible to activate or deac-
tivate the lock function.
c.)
You can activate (Lock
On
) or deactivate (Lock
Off
) the lock function by pressing the right turn
button.
The activated lock function is displayed by the = icon in the
right display. If you cannot see the = icon in the right dis-
play, the lock function of the agitation table is turned off.
Setting the agitation time