HS/KS260 0305
29
place evenly on the agitation table, and that all vessels are secu-
rely fastened in place.
AS 260.1
Universal attachment
AS 260.5
Separating funnel attachment
Basic error codes:
ATTENTION!
If an error code is encountered, an audio warning
signal is generated and the red Off signal light goes on in the dis-
play field. First, please try to continue operation by turning the
device off and back on again.If it is still not possible to eliminate
an error after an extended wait, please contact our service
department. If you do so, you should always tell us what error
code has been encountered. This simplifies the process of
tracking down the error and makes it possible to form a prelimi-
nary opinion.
Error code
Error
Cause of the error
Off - LED (red)
The potentiometer for
-Too rapid a change in the
1x flash /
the speed is not assig-
speed target value
2 sec pause
ning any target value
-Internal device error
Off - LED (red)
Permissible motor tem- -Permissible supported
2x flash /
peratur exceeded
weight exceeded
2 sec pause
-Permissible ambient
temperature exceeded
Motor locked
-Agitation table is being
hindered in this stroke mo-
tion by an outside effect
Read fork signal not
-Internal device error
available
Off - LED (red)
Maximum adjustable
-Resonance behavior of
3x flash /
speed of the device
the experiment setup
2 sec pause
exceeded
Off - LED (red)
Motor standstill for de-
-When turning on the
4x flash /
vice switch-on proce-
device, you must ensure
2 sec pause
dure
(safety confirma-
that the motor is not
tion not ensured
performing any rational
motion
Off - LED (red)
Safety relay cannot
-Internal device error
5x flash /
be activated
2 sec pause
Off - LED (red)
The switch for the TIME -Internal device error
6x flash /
function is not assigning
2 sec pause
any target value
Off - LED (red)
Problem with the power -The device is beeing
7x flash /
supply
operating with under vol-
2 sec pause
tage or overvoltage (for
permissible voltage range,
see technical data)
HS 260
KS 260
1/min
g
0
100
150
200
250
300
350
400
450
500
50
0
8000
7000
6000
5000
4000
3000
2000
1000
g
1/min
HS 260
KS 260
0
100
150
200
250
300
350
400
450
500
50
0
8000
7000
6000
5000
4000
3000
2000
1000
Error codes
3106600c_HSKS260_052006:HS-KS 260 b-c 13.04.10 14:32 Seite 29