Speaking, moving arm or body, being

angry, excited or nervous during test

Retest when calm; avoid speaking or movement during

the test

Bluetooth

connection

unstable

Bluetooth connection unsuccessful,

monitor is abnormal, or strong

electromagnetic interference is present

Reset iOS/Android device. Reset monitor by pressing

the

button . Make sure the monitor and

iOS/Android device are away from other electrical

equipment. Please see GENERAL SAFETY AND

PRECAUTIONS

No response

Incorrect

operation

or

strong

electromagnetic interference

Press the

button to reset the device and reconnect

the iOS/Android device to the monitor, relaunch app.

CARE AND MAINTENANCE

1. If this monitor is stored near freezing temperatures, allow it to return to room temperature before use.

2. If the monitor is not used for a long time, please be sure to fully charge it every month.

3. It is recommended that product performance be checked every 2 years or after each repair. Please contact the iHealth

Customer Service Center to do so.

4. No monitor component needs to be maintained by the user. The circuit diagrams, component part lists, descriptions,

calibration instructions, or other information which will assist the user’s appropriately qualified technical personnel to repair

those parts of the equipment which are designated for repair can be supplied by the iHealth technical department.

5. Clean the monitor with a dry, soft cloth or a moistened and well wrung soft cloth using water, diluted disinfectant alcohol, or

diluted detergent.

6. The monitor can maintain the safety and performance characteristics for a minimum of 10,000 measurements or three

years of usage, and the cuff integrity is maintained after 1,000 open close cycles.

7. The battery can maintain the performance characteristics for a minimum of 300 charge cycles.

8. It is recommended that if the cuff is used in a hospital or a clinic, it be disinfected twice a week. Wipe the inner side (the side

that contacts skin) of the cuff with a soft cloth lightly moistened with Ethyl alcohol (75-90%). Then air dry the cuff.

Do not drop this monitor or subject it to strong impact.

Avoid high temperature and direct sunlight. Do not immerse the monitor in water as this will result in damage to the

monitor.

Do not attempt to disassemble this monitor.

Battery replacement should only be performed by a qualified iHealth technician. To do otherwise will void your warranty

and possibly damage your unit.

Cuff replacement should only be performed by a qualified iHealth technician. To do otherwise will possibly damage your

unit.

WARRANTY INFORMATION

The iHealth Wireless Blood Pressure Monitor is warranted to be free from defects in materials and workmanship within one

year from the date of purchase when used in accordance with the provided instructions. The warranty extends only to the end

user. We will, at our option, repair or replace without charge the iHealth Wireless Blood Pressure Monitor covered by the

warranty. Repair or replacement is our only responsibility and your only remedy under the warranty.

EXPLANATION OF SYMBOLS

Symbol for "TYPE BF APPLIED PARTS" (cuff only)

Symbol for "THE OPERATION GUIDE MUST BE READ"

The sign background color: blue The sign graphical symbol: white

Summary of Contents for BP3

Page 1: ...SET UP REQUIREMENTS 2 SET UP PROCEDURES 2 SPECIFICATIONS 4 GENERAL SAFETY AND PRECAUTIONS 4 BATTERY HANDLING AND USAGE 4 TROUBLESHOOTING 4 CARE AND MAINTENANCE 4 WARRANTY INFORMATION 4 EXPLANATION OF...

Page 2: ...using a technique in which an inflatable cuff is wrapped around the upper arm The measurement range of the standard cuff circumference is 8 21 32 to 14 3 16 22cm 36cm Note Consult your physician for...

Page 3: ...or Status Bluetooth Indicator Waiting to connect No light Connected and measuring Steady green light Measurement completed and Device far away from the monitor Extinguishing green light MEASUREMENT PR...

Page 4: ...y remove it from the monitor and clean it with a moistened cloth Do not rinse the monitor or cuff with running water Press the button at any time to interrupt a measurement Note Physical activity eati...

Page 5: ...one call comes in during the measurement the measurement process will be terminated automatically It is thus recommended that the phone be set in Airplane mode during measurement to avoid interrupting...

Page 6: ...dicator Charging Flashing green light Fully charged Steady green light Low battery Flashing red light for a few seconds Abnormal state Steady red light Do not change the battery If the battery can no...

Page 7: ...gent 6 The monitor can maintain the safety and performance characteristics for a minimum of 10 000 measurements or three years of usage and the cuff integrity is maintained after 1 000 open close cycl...

Page 8: ...andards Apple is not responsible for the operation of this device or its compliance with safety and regulatory standards Please note that the use of this accessory with iPod iPhone or iPad may affect...

Page 9: ...conditi ons 1 this device may not cause interference and 2 this device must accept any interference including interference that may cause undesired operation of the device This product is approved in...

Page 10: ...ustomer or the user of the BP3L should assure that it is used in such an environment IMMUNITY test IEC 60601test level Compliance level Electromagnetic environment guidance Electrostatic discharge ESD...

Page 11: ...o any part of the BP3L including cables than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter Recommended separation distance 80 MHz to 8...

Page 12: ...by maintaining a minimum distance between portable and mobile RF communications equipment transmitters and the BP3L as recommended below according to the maximum output power of the communications equ...