ELECTROMAGNETIC COMPATIBILITY INFORMATION

Table 1

For all ME EQUIPMENT and ME SYSTEMS

Guidance and manufacture’s declaration - electromagnetic emissions

The BP3L is intended for use in the electromagnetic environment specified below.

The customer or the user of the BP3L should assure that it is used in such an environment.

Emissions test

Compliance

Electromagnetic environment - guidance

RF emissions

CISPR 11

Group 1

The BP3L uses RF energy only for its internal function. Therefore, its RF

emissions are very low and are not likely to cause any interference in

nearby electronic equipment.

RF emissions

CISPR 11

Class B

The BP3L is suitable for use in all establishments, including domestic

establishments and those directly connected to the public low-voltage

power supply network that supplies buildings used for domestic

purposes.

Harmonic emissions

IEC 61000-3-2

Class A

Voltage fluctuations/

flicker emissions

IEC 61000-3-3

Complies

Table 2

For all ME EQUIPMENT and ME SYSTEMS

Guidance and manufacturer’s declaration - electromagnetic immunity

The BP3L is intended for use in the electromagnetic environment specified below. The customer or the user of the BP3L

should assure that it is used in such an environment.

IMMUNITY test

IEC 60601test level

Compliance level

Electromagnetic environment - guidance

Electrostatic discharge

(ESD)

IEC 61000-4-2

± 6 kV contact

± 8 kV air

± 6 kV contact

± 8 kV air

Floors should be wood, concrete or ceramic

tile. If floors are covered with synthetic

material, the relative humidity should be at

least 30 %.

Electrical fast

transient/burst

IEC 61000-4-4

± 2 kV for power supply

lines

± 1 kV for input/output

lines

± 2 kV for power

supply lines

Mains power quality should be that of a

typical commercial or hospital environment.

Surge

IEC 61000-4-5

± 1 kV line(s) to line(s)

± 2 kV line(s) to earth

± 1 kV line(s) to

line(s)

± 2 kV line(s) to

earth

Mains power quality should be that of a

typical commercial or hospital environment.

Voltage dips, short

interruptions and

voltage variations

on power supply

input lines

IEC 61000-4-11

<5 % U

T

(>95 % dip in U

T

)

for 0.5 cycle

40 % U

T

(60 % dip in U

T

)

for 5 cycles

70 % U

T

(30 % dip in U

T

)

<5 % U

T

(>95 % dip in U

T

)

for 0.5 cycle

40 % U

T

(60 % dip in U

T

)

for 5 cycles

70 % U

T

(30 % dip in U

T

)

Mains power quality should be that of a

typical commercial or hospital environment. If

the user of the BP3L requires continued

operation during power mains interruptions, it

is recommended that the BP3L be powered

from an uninterruptible power supply or a

battery.

Summary of Contents for BP3

Page 1: ...SET UP REQUIREMENTS 2 SET UP PROCEDURES 2 SPECIFICATIONS 4 GENERAL SAFETY AND PRECAUTIONS 4 BATTERY HANDLING AND USAGE 4 TROUBLESHOOTING 4 CARE AND MAINTENANCE 4 WARRANTY INFORMATION 4 EXPLANATION OF...

Page 2: ...using a technique in which an inflatable cuff is wrapped around the upper arm The measurement range of the standard cuff circumference is 8 21 32 to 14 3 16 22cm 36cm Note Consult your physician for...

Page 3: ...or Status Bluetooth Indicator Waiting to connect No light Connected and measuring Steady green light Measurement completed and Device far away from the monitor Extinguishing green light MEASUREMENT PR...

Page 4: ...y remove it from the monitor and clean it with a moistened cloth Do not rinse the monitor or cuff with running water Press the button at any time to interrupt a measurement Note Physical activity eati...

Page 5: ...one call comes in during the measurement the measurement process will be terminated automatically It is thus recommended that the phone be set in Airplane mode during measurement to avoid interrupting...

Page 6: ...dicator Charging Flashing green light Fully charged Steady green light Low battery Flashing red light for a few seconds Abnormal state Steady red light Do not change the battery If the battery can no...

Page 7: ...gent 6 The monitor can maintain the safety and performance characteristics for a minimum of 10 000 measurements or three years of usage and the cuff integrity is maintained after 1 000 open close cycl...

Page 8: ...andards Apple is not responsible for the operation of this device or its compliance with safety and regulatory standards Please note that the use of this accessory with iPod iPhone or iPad may affect...

Page 9: ...conditi ons 1 this device may not cause interference and 2 this device must accept any interference including interference that may cause undesired operation of the device This product is approved in...

Page 10: ...ustomer or the user of the BP3L should assure that it is used in such an environment IMMUNITY test IEC 60601test level Compliance level Electromagnetic environment guidance Electrostatic discharge ESD...

Page 11: ...o any part of the BP3L including cables than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter Recommended separation distance 80 MHz to 8...

Page 12: ...by maintaining a minimum distance between portable and mobile RF communications equipment transmitters and the BP3L as recommended below according to the maximum output power of the communications equ...

Reviews: