background image

electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the

measured field strength in the location in which the BP3L is used exceeds the applicable RF compliance level above, the

BP3L should be observed to verify normal operation. If abnormal performance is observed, additional measures may be

necessary, such as re-orienting or relocating the BP3L.

b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3V/m.

Table 4

For ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING

Recommended separation distances between portable and mobile RF communications equipment and the BP3L

The BP3L is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The

customer or the user of the BP3L can help prevent electromagnetic interference by maintaining a minimum distance between

portable and mobile RF communications equipment (transmitters) and the BP3L as recommended below, according to the

maximum output power of the communications equipment.

Rated maximum output

power of transmitter

W

Separation distance according to frequency of transmitter

m

150 kHz to 80 MHz

P

d

2

.

1

=

80 MHz to 800 MHz

P

d

2

.

1

=

800 MHz to 2,5 GHz

P

d

3

.

2

=

0,01

0,12

0,12

0,23

0,1

0,38

0,38

0,73

1

1,2

1,2

2,3

10

3,8

3,8

7,3

100

12

12

23

For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m)

can be determined using the equation applicable to the frequency of the transmitter, where P is the maximum output power

rating of the transmitter in watts (W) according to the transmitter manufacturer.

NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.

NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection

from structures, objects and people.

BP3L-SMSY01 V1.0

Summary of Contents for BP3

Page 1: ...SET UP REQUIREMENTS 2 SET UP PROCEDURES 2 SPECIFICATIONS 4 GENERAL SAFETY AND PRECAUTIONS 4 BATTERY HANDLING AND USAGE 4 TROUBLESHOOTING 4 CARE AND MAINTENANCE 4 WARRANTY INFORMATION 4 EXPLANATION OF...

Page 2: ...using a technique in which an inflatable cuff is wrapped around the upper arm The measurement range of the standard cuff circumference is 8 21 32 to 14 3 16 22cm 36cm Note Consult your physician for...

Page 3: ...or Status Bluetooth Indicator Waiting to connect No light Connected and measuring Steady green light Measurement completed and Device far away from the monitor Extinguishing green light MEASUREMENT PR...

Page 4: ...y remove it from the monitor and clean it with a moistened cloth Do not rinse the monitor or cuff with running water Press the button at any time to interrupt a measurement Note Physical activity eati...

Page 5: ...one call comes in during the measurement the measurement process will be terminated automatically It is thus recommended that the phone be set in Airplane mode during measurement to avoid interrupting...

Page 6: ...dicator Charging Flashing green light Fully charged Steady green light Low battery Flashing red light for a few seconds Abnormal state Steady red light Do not change the battery If the battery can no...

Page 7: ...gent 6 The monitor can maintain the safety and performance characteristics for a minimum of 10 000 measurements or three years of usage and the cuff integrity is maintained after 1 000 open close cycl...

Page 8: ...andards Apple is not responsible for the operation of this device or its compliance with safety and regulatory standards Please note that the use of this accessory with iPod iPhone or iPad may affect...

Page 9: ...conditi ons 1 this device may not cause interference and 2 this device must accept any interference including interference that may cause undesired operation of the device This product is approved in...

Page 10: ...ustomer or the user of the BP3L should assure that it is used in such an environment IMMUNITY test IEC 60601test level Compliance level Electromagnetic environment guidance Electrostatic discharge ESD...

Page 11: ...o any part of the BP3L including cables than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter Recommended separation distance 80 MHz to 8...

Page 12: ...by maintaining a minimum distance between portable and mobile RF communications equipment transmitters and the BP3L as recommended below according to the maximum output power of the communications equ...