Whether applying the device to confused, dazed, unconscious or otherwise incapable patients need to consult a 

physician. The device should be used only under supervision if the physician agrees. 

z

 

Whether applying the device to the patients with serious mobility impairments need to consult a physician. The 

device should be used only under supervision if the physician agrees. 

z

 

The device should not be exposed to strong electromagnetic fields, otherwise they may cause inaccurate results 

or malfunction. 

z

 

Do not use the device in combination with high-frequency(HF) surgical equipment. 

z

 

Do not use the device in the presence of flammable anaesthetics due to risk of explosion. 

z

 

The physician must be certain that, according to the health of the patient, the use of the device will not damage 

blood circulation in the arm. 

z

 

Do not attach the cuff to a limb being used for intravenous infusions. This may cause the infusion to be blocked 

and cause the patient harm. 

z

 

The cuff must not be placed on the area with dermatitis, open wounds and other contraindications. Please 

consult a physician for specific situation. 

Caution 

z

 

Motion, trembling, shivering may affect the measurement reading. 

z

 

Blood pressure measurements determined by this product are equivalent to those obtained by professional 

healthcare practitioners using the cuff/stethoscope auscultation method within the limits prescribed by the 

American National Standard, Electronic or Automated Sphygmomanometer. 

z

 

For information regarding potential electromagnetic or other interference between the blood pressure monitor 

and other devices together with advice regarding avoidance of such interference, please see 

ELECTROMAGNETIC COMPATIBILITY INFORMATION. It is suggested that the blood pressure monitor be kept 

10 meters away from other wireless devices, such as WLAN unit, microwave oven, etc. 

z

 

The results of blood pressure measurements may be influenced by: the patient's physical condition, cuff 

wrapping method, measuring posture, measuring environment etc. Strictly in accordance with product 

instructions to operate the device, otherwise it may lead to inaccurate results. 

z

 

It is recommended to apply the cuff to non-dominant arm. Choose the arm of the other side if there are blood 

pressure measurement contraindications(e.g. there are open wounds or amputation on the area where the cuff 

is wrapped) on the arm of one side. Please consult a physician for specific situation. 

z

 

Choose the appropriate cuff according to the arm circumference. The arm circumference outside the applicable 

scope of the cuff may lead to inaccurate results. 

z

 

Cuff wrapping method should be strictly in accordance with product instructions. Inappropriate cuff wrapping 

location, cuff wrapping too loose or too tight, and other factors may lead to inaccurate results. 

z

 

When the cuff is placed over thick clothing may result in inaccurate readings. 

z

 

Relax and be quiet during measurement, avoid talking and movement. Avoid flexing the muscles or moving the 

hand and fingers of the cuffed arm during the measurement. 

z

 

Do not use the device when it is exposed to mechanical vibration (e.g. in vehicles). Advise patient not to drive 

but if this is necessary to stop if possible during measurement. 

z

 

Please check measurement values by other methods (e.g. auscultation), if you suspect an value. 

z

 

To prevent electric shock hazard due to leakage current, only use the power supplies which are compliant with 

the technical specifications of the device. 

z

 

The equipment must never be connected to a printer, computer or other external equipment while still fitted to 

the patient. 

Summary of Contents for ABP100

Page 1: ...NTS 6 BATTERY HANDLING AND USAGE 6 SET UP PROCEDURES 7 ABPM SYNC AND SETUP 7 MEASUREMENT PROCEDURES 8 TAKING YOUR BLOOD PRESSURE READING 9 TAKING YOUR BLOOD PRESSURE READING MANUALLY 9 TAKING YOUR BLO...

Page 2: ...hypertension z Assessment of treatment Assessing 24 h BP control Identifying true resistant hypertension z Assessing morning hypertension and morning BP surge z Screening and follow up of obstructive...

Page 3: ...ce of adjacent pulse period is 0 14s and more than 53 percent of the total number of pulses readings falls within this definition z This product should not be used as a USB device z If the blood press...

Page 4: ...IBILITY INFORMATION It is suggested that the blood pressure monitor be kept 10 meters away from other wireless devices such as WLAN unit microwave oven etc z The results of blood pressure measurements...

Page 5: ...l of the cuff Please consult a physician for specific situation z It is recommended that the patient record important events such as take medicine taking time drug name drug dose etc awake and asleep...

Page 6: ...asurements on a full charge 19 The blood pressure measurement system includes accessories pump valve cuff LED screen and sensor Note These specifications are subject to change without notice SET UP RE...

Page 7: ...Health service provider Battery life and charge cycles vary by use and settings SET UP PROCEDURES Download the Free iHealth App Prior to first use scan QR code and download the iHealth App Follow the...

Page 8: ...surement Lying Down During Measurement a Lie on your back b Place your arm straight along your side with your hand palm side up c The cuff should be placed at the same level as your heart Note Blood p...

Page 9: ...tep to achieving a successful test TAKING YOUR BLOOD PRESSURE READING MANUALLY a Long press the button the monitor will activate and all display characters are shown for self test You can check the LE...

Page 10: ...d When Auto Mode on the monitor will continue take measurements in the scheduled time and after the monitor get enough readings the monitor automatic stop taking readings Note Please consult a health...

Page 11: ...g the button and holding for about 10 seconds then connect the mobile device correctly and try again If the monitor is still abnormal please contact the local distributor or the factory LED display re...

Page 12: ...2 years or after each repair Please contact the customer service z Do not use the device if you think it is damaged or if anything appears unusual Please refer your device to qualified service personn...

Page 13: ...Paris France support ihealthlabs eu www ihealthlabs eu ANDON HEALTH CO LTD No 3 Jinping Street YaAn Road Nankai District Tianjin 300190 China Tel 86 22 60526161 IMPORTANT INFORMATION REQUIRED BY THE...

Page 14: ...pe and its gain should be so chosen that the equivalent isotropically radiated power e i r p is not more than that necessary for successful communication Conform ment la r glementation d Industrie Can...

Page 15: ...n IEC 61000 3 2 Class A Home healthcare environment Voltage fluctuations and flicker IEC 61000 3 3 Compliance Home healthcare environment Table 2 Enclosure Port Phenomenon Basic EMC standard Immunity...

Page 16: ...4 2 kV 100kHz repetition frequency Surges Line to line IEC 61000 4 5 0 5 kV 1 kV Surges Line to ground IEC 61000 4 5 0 5 kV 1 kV 2 kV Conducted disturbances induced by RF fields IEC 61000 4 6 3V 0 15...

Reviews: