16

 

/

 

16

 

 

870 
930 
1720 

1700-1990 

Pulse modulation 217Hz, 28V/m 

1845 
1970 
2450 

2400-2570 

Pulse modulation 217Hz, 28V/m 

5240 

5100-5800 

Pulse modulation 217Hz, 9V/m 

5500 
5785 

Table 4 – Input a.c. power Port

 

Phenomenon Basic EMC 

standard 

Immunity test levels 

Home Healthcare Environment 

Electrical fast 
transients/burst 

IEC 61000-4-4 
 

±2 kV 
100kHz repetition frequency 

Surges 
Line-to-line 

IEC 61000-4-5 
 

±0.5 kV, ±1 kV 
 

Surges 
Line-to-ground 

IEC 61000-4-5 
 

±0.5 kV, ±1 kV, ±2 kV 
 

Conducted 
disturbances 
induced by RF fields 

IEC 61000-4-6 

3V, 0.15MHz-80MHz 
6V in ISM and amateur radio bands between 
0.15MHz and 80MHz 
80%AM at 1kHz 

Voltage dips 

IEC 61000-4-11 

0%    U

T

; 0.5 cycle 

At 0º, 45º,    90º, 135º, 180º, 225º, 270º and 315º 
0% U

T

; 1 cycle 

and 
70% U

T

; 25/30 cycles 

Single phase: at 0º 

Voltage interruptions  IEC 61000-4-11 

0% U

T; 

250/300 cycles 

 

Date of issue:Sep25,2017 
ABP100-SMSY01 V1.0

 

Summary of Contents for ABP100

Page 1: ...NTS 6 BATTERY HANDLING AND USAGE 6 SET UP PROCEDURES 7 ABPM SYNC AND SETUP 7 MEASUREMENT PROCEDURES 8 TAKING YOUR BLOOD PRESSURE READING 9 TAKING YOUR BLOOD PRESSURE READING MANUALLY 9 TAKING YOUR BLO...

Page 2: ...hypertension z Assessment of treatment Assessing 24 h BP control Identifying true resistant hypertension z Assessing morning hypertension and morning BP surge z Screening and follow up of obstructive...

Page 3: ...ce of adjacent pulse period is 0 14s and more than 53 percent of the total number of pulses readings falls within this definition z This product should not be used as a USB device z If the blood press...

Page 4: ...IBILITY INFORMATION It is suggested that the blood pressure monitor be kept 10 meters away from other wireless devices such as WLAN unit microwave oven etc z The results of blood pressure measurements...

Page 5: ...l of the cuff Please consult a physician for specific situation z It is recommended that the patient record important events such as take medicine taking time drug name drug dose etc awake and asleep...

Page 6: ...asurements on a full charge 19 The blood pressure measurement system includes accessories pump valve cuff LED screen and sensor Note These specifications are subject to change without notice SET UP RE...

Page 7: ...Health service provider Battery life and charge cycles vary by use and settings SET UP PROCEDURES Download the Free iHealth App Prior to first use scan QR code and download the iHealth App Follow the...

Page 8: ...surement Lying Down During Measurement a Lie on your back b Place your arm straight along your side with your hand palm side up c The cuff should be placed at the same level as your heart Note Blood p...

Page 9: ...tep to achieving a successful test TAKING YOUR BLOOD PRESSURE READING MANUALLY a Long press the button the monitor will activate and all display characters are shown for self test You can check the LE...

Page 10: ...d When Auto Mode on the monitor will continue take measurements in the scheduled time and after the monitor get enough readings the monitor automatic stop taking readings Note Please consult a health...

Page 11: ...g the button and holding for about 10 seconds then connect the mobile device correctly and try again If the monitor is still abnormal please contact the local distributor or the factory LED display re...

Page 12: ...2 years or after each repair Please contact the customer service z Do not use the device if you think it is damaged or if anything appears unusual Please refer your device to qualified service personn...

Page 13: ...Paris France support ihealthlabs eu www ihealthlabs eu ANDON HEALTH CO LTD No 3 Jinping Street YaAn Road Nankai District Tianjin 300190 China Tel 86 22 60526161 IMPORTANT INFORMATION REQUIRED BY THE...

Page 14: ...pe and its gain should be so chosen that the equivalent isotropically radiated power e i r p is not more than that necessary for successful communication Conform ment la r glementation d Industrie Can...

Page 15: ...n IEC 61000 3 2 Class A Home healthcare environment Voltage fluctuations and flicker IEC 61000 3 3 Compliance Home healthcare environment Table 2 Enclosure Port Phenomenon Basic EMC standard Immunity...

Page 16: ...4 2 kV 100kHz repetition frequency Surges Line to line IEC 61000 4 5 0 5 kV 1 kV Surges Line to ground IEC 61000 4 5 0 5 kV 1 kV 2 kV Conducted disturbances induced by RF fields IEC 61000 4 6 3V 0 15...

Reviews: