REMOTE CHANGE
(I/O PORT1 10PIN)
Remote change
10PIN
C o n t r o l A c t i o n
0
EXIT I/O MODE
1
Change EXIT I/O Mode to RS232C Mode
CONTROL SIGNAL
(I/O PORT1 6
〜
8PIN,19,20PIN)
9PIN 26PIN 28PIN
F U N C T I O N
0
0
0
No Control signal.
1
0
0
Reset alarm.
Input this signal to reset when alarm/error signal shows
a number ‘1’. Empty alarm is also reset.
Error screen and/or buzzer sound is reset.
0
0
1
Reset settings.
Reset compensating condition to an initial setting.
Reset dispensing pressure and vacuum pressure to the
initial settings.
Reset a number of dispensing
Reset the lapsed time during viscosity compensation in
use..
During alarm/error, this signal is ignored.
0
1
0
Reset.
After input this signal, the software of controller is reset
and restarts from the initial title screen.
During ‘Busy’, this signal is ignored.
Remark
:
If the input cable is not connected in the control signal, the controller is regarded as no signal.