ACCURA9/9L INSTRUCTION MANUAL
Set the barrel filled up to the upper limit when the vacuum condition is downloaded.
Down-load of Vacuum Conditions
L C D S C R E E N
P R O C E D U R E S
1
D.Samp Function Setting Screen
*** D.Samp Menu ***
Ch 00 Bot̲Level NOD
Top̲Level NOD
Emp̲Level OFF
Down load the vacuum condition data:
Shift the screen to D.SAMP function setting.
Check that Bot_Lovel displays NOD at D.SAMP
function setting screen.
Put the cursor on the Top_Level.
Press the
CTRL+RETURN
key to shift the next
screen.
2
Barrel Confirmation Screen
BARREL(Top) READY?
YES .. RETURN
NO .. CLR RESET
Set the barrel filled materials up to the upper level
and press
RETURN key.
Press
CLR RESET key
to stop.
3
Top Sampling Screen
*** Top Level ***
Ch 00
Vc.Press 015.0
Vt 00.500 Vi 01.000
Input the vacuum pressure (Vc.Press), the vacuum
time (Vt.) and the vacuum interval time (Vi) with
/OFF
key while keeping off material dropping and
bubbling.
Input more than 30msec. at the
Vt
、
Vi
.
If O.K., press
RETURN key
.
Press
SHIFT+CLR RESET
key to stop.
4
D.sanp Function Setting Screen
*** D.Samp Menu ***
Ch 00 Bot̲Level NOD
Top̲Level NOD
Emp̲Level OFF
Remain Top_Level being NOD at D.SAMP function
setting screen.