Procedure • Disinfect Handheld Device, Base Station and connecting
cables by wiping off the surfaces with a lint free cloth moist-
ened with disinfectant.
• Disinfect the respiratory components in a disinfection bath
consulting the disinfectant instructions.
• Rinse the breathing bag with the disinfectant.
• Make sure that all parts can dry properly.
• Dry the breathing bag with opening downwards.
• See also chapter “Drying”, page 23 and “Storage”, page 23.
Disinfection
by boiling
Boiling may be used if the respiratory components get used by
only one person. The respiratory components have to be steril-
ized if they are used by several persons (see page 29).
Disinfection
by boiling
Boil with
opening
upwards
Dry with opening
downwards
Procedure • Make sure the parts are fully covered by water and do not
contain air.
WARNING
Hot water!
Risk of burns.
• Fill the breathing bag with hot water before boiling it.
• Make sure that no air remains inside the breathing bag (open-
ing upwards).
• Make sure that the breathing bag is kept open and is not
folded during boiling.
• Start counting the duration when the water reaches a rolling
boil.
• Duration: 8-10 minutes
• Make sure that all parts can dry properly.
• Dry the breathing bag with opening downwards.
• See also chapter “Drying”, page 23 and “Storage”, page 23.