Instruction for Use

Portable Mesh

Nebulizer

NEB-01-C

Summary of Contents for NEB-01-C

Page 1: ...Instruction for Use Portable Mesh Nebulizer NEB 01 C...

Page 2: ...Instruction for Use 5 Cleaning and Disinfection Method 6 Technical Data 7 Trouble Shooting Guideline 8 Declaration of Conformity 9 Warranty 4 1 Before First Use 4 2 Installation 4 3 Charging 4 4 Inst...

Page 3: ...d by persons with reduced physical sensory or mental capabilities or by persons with little or no knowledge of the appliance as long as they are under proper supervision and have received instructions...

Page 4: ...ck or malfunction Do not touch the center of spray mesh by hand or other sharp objects It may cause damage so you will not be able to use it The nebulizer has no moisture and dust function and the pro...

Page 5: ...erous and prohibited For your own safety don t remove the battery from the product and please contact only authorized Zepter Service Centre 5 3 Item Description 3 1 Content of Packaging 3 2 Device des...

Page 6: ...apter with rated output DC 5 0 V 1A 3 While the device is charging the green indicator ashes continually 4 The LED indicator is lit steadily when the battery is fully charged 5 The appliance can be us...

Page 7: ...ight ashes ten times While the orange light is ashing press switch key once to continue work Long press for 3 seconds to switch the atomization mode Blue light keeps on uniform nebulization mode Blue...

Page 8: ...an t be shared to avoid cross infection 4 5 Nebulization 1 Connect mask or mouthpiece click to switch power on start nebulization 2 It starts with a medium spray rate and spray rate can be adjusted by...

Page 9: ...l liquid in the medication cup and do not reuse 9 Use clean water to clean the medication cup spray nozzle lid threaded assemblies and accessories and then follow the recommended method of disinfectio...

Page 10: ...clean it with wet sterile medical gauze After cleaning keep the main unit dry 4 Store all the parts in a dry and clean place to avoid contamination 10 2 Disinfection After each use it is necessary to...

Page 11: ...r disinfection rinse all the parts with distilled water then wipe with clean and sterile medical gauze or air dry naturally to keep all parts dry Notes Disinfectants remaining on components need to be...

Page 12: ...4 The spray mesh may be blocked you can drop 2 or 3 drops of white vinegar into the medication cup with 3 6 ml of water Clean the medication cup to disinfect for next nebulization 1 Please refer to a...

Page 13: ...er cannot be held responsible for any damage deriving from improper incorrect or unreasonable use All our appliances have undergone strict inspection test and are guaranteed against manufacturing defe...

Page 14: ...hts under applicable national legislation in force nor the customer s rights against the retailer arising from the sales purchase agreement In the absence of applicable national legislation this guara...