22
Maintenance
Maintenance menu
WARNING
The Base Station menu “maintenance” may only be used by
authorized service personal.
Inspection by the user
A close inspection of the ST Medical
®
is necessary before each
therapy session. If you suspect that the ST Medical
®
may be mal-
functioning, send it to your local distributor to be examined and/
or repaired. Necessary repair work should only be carried out by
Idiag or an authorized customer service center. Faulty compo-
nents which may affect the safety of the ST Medical
®
equipment
must be replaced by genuine spare parts.
WARNING
• Clean and disinfect or clean and sterilize the respiratory
components before the first use in accordance with chapter
‚maintenance‘. Pages 22-29.
• Make an inspection of the ST Medical
®
before each therapy
session.
• Do not use a ST Medical
®
which has not passed an inspection.
• Do not use a ST Medical
®
that is damaged or does not func-
tion properly.
• Do not use a ST Medical
®
if a battery leaks.
• Do not open the Handheld Device or the Base Station (except
the battery compartment).
Interval
Scope
Method
Before each
therapy
• Perform a function control
(see page 14).
Inspection
• Make sure that the piston moves friction-
less.
Inspection
• Make sure that the ST Medical
®
is in hy-
gienically perfect condition. See chapter‚
Maintenance‘ pages
23-29.
Inspection
• Make sure that breathing bag and mouth-
piece have no leak parts or cracks.
Inspection
• Look for damaged components, labels and
warning signs.
Inspection
• Check all cables (damage, breakage).
Inspection
• Make sure that nobody can be injured with
the ST Medical
®
.
Inspection