13
Setup the Base Station
2 Batteries
Dummy
removable
4 Batteries
1.
Insert the batteries as shown.
The battery compartment of the ST Medical
®
holds up to
4 batteries. (The ST Medical
®
may also run on 2 batteries,
which reduces their life span accordingly.)
The following situations require the installa-
tion of new batteries:
please change
battery
• before you first use the ST Medical
®
• if “please change battery” is displayed
• when the battery power is too low to run the ST Medical
®
WARNING
• Open the cover of the battery compartment on the rear side to
insert the batteries as shown, making sure to align the positive
and the negative poles correctly to avoid damage to the unit.
• To extend battery life, remove the batteries if ST Medical
®
is not
likely to be used for several days or weeks.
• Note: Batteries do contain corrosive liquid. If a battery leaks:
Do not expose skin, eyes, or mouth to the liquid. If you do,
flush the affected body parts with water for at least 15 minutes.
Consult a doctor. Do not inhale vapor. Immediately remove any
liquid from the unit with a damp cloth and dispose of the cloth.
Idiag accepts no responsibility for injuries to customers, or
damage to the device, caused by leaking or defective batteries.
• Do not recharge non-rechargeable batteries.
• Take care of our environment; properly dispose of used batter-
ies!
• Battery life: see ‚Technical Data‘
• Always use high quality batteries. See ‚Technical Data‘
2.
Perform a function check. See page 14.