Operator Manual Aspect 100UC Plus
L100-3065 REV C
USER REPLACEMENT ITEMS
These are easily replaced by the operator.
•
Tabletop Pad
Pads are held in place by hook and loop fasteners. Simply pull pad up gently to
remove. New replacement pads are supplied with new self-adhesive fasteners
installed and new mating pieces for the tabletop or extension. Peel off old
fasteners from tabletop. Peel off the protective paper and carefully place new
pad into position. Apply pressure to complete installation.
Note: Pads comply to California Technical Bulletin 117 as well as Underwriters
Laboratories 94 (UL94). The vinyl coating is self-extinguishing when free of
solvents.
LATEX
Not made with natural rubber latex
Tabletop Pad
2in. x 28in. x 46-1/2in. (5cm x 71cm x 118cm)
Part Number X100-1818
Extension Pad
2in. x 24in. x 34in. (5cm x 61cm x 86cm)
Part Number X100-1817
TROUBLE-SHOOTING
SYMPTOM
POSSIBLE CAUSE
ACTION
No motor sound
Power cord not plugged in
Connect power cord
And/or movement
No power at the receptacle
Check appropriate
circuit breaker
A missing or loose connection
Inspect wires and
at the control box
connections
A blown fuse
Change fuse with
same rating
A damaged wire
Repair or replace wire
Emergency stop button activated
Twist & pull up
For other symptoms, consult field service or linak website for actuator (motor)
and controller
.
For other symptoms, consult field service.