Operator Manual Aspect 100UC Plus
L100-3065 REV C
•
HAND CONTROL
•
First of buttons moves tabletop up and down
•
Second row of buttons controls the Trendelenburg motion
•
Third row of buttons slides the tabletop longitudinally
•
Fourth row of buttons slides the tabletop latterly
To move tabletop, press appropriate button and hold button depressed until desired position is
reached.
The tabletop will pause at the horizontally level position during Trendelenburg motion. Release and
activate control again to continue past level. As the position of the patient’s head may be at either end
of the table, the action may be opposite to what is indicated by the label on the control.
Hand Control
Part Number K000-0158 plus label D100-1563
CROSS EXT.
S LIDE BACK
REV . TREN
HEIGHT UP
CROS S RET.
S LIDE FOWARD
TREN
HEIG HT DOWN
CROSS EXT.
SLIDE BACK
TREN
HEIGHT UP
CROSS RET.
SLIDE FOWARD
REV. TREN
HEIGHT DOWN
CAUTION
Always check that the patient is secure before activating
any table motion