Operator Manual
Aspect 100-4T G3 and ISR
G3
L100-2843 28-MAR-2018
6
5.
EMC (Electromagnetic Compatibility) STATEMENT
Portable and Mobile RF Communications Equipment can affect Medical Electrical
Equipment including the
Aspect
100-4T G3 and
Aspect
ISR G3 tables. Use special precautions
regarding EMC when these tables are installed, operated and maintained. EMC operating
specifications for these tables are in the SPECIFICATIONS section of this manual (Section 16).
The use of accessories, transducers and/or cables other than those specified, with the
exception of those sold by the manufacturer as replacement parts for internal components, may
result in increased emissions or decreased immunity of the equipment or system.
WARNING: The
Aspect
100-4T G3 or
Aspect
ISR G3 should not be used
adjacent to or stacked with other medical electrical equipment and that if adjacent
or stacked use is necessary, the equipment or system should be observed to verify
normal operation in the configuration in which it will be used.
WARNING: Portable RF communications equipment (including peripherals
such as antenna cables and external antennas) should be used no closer than 30
cm (12 inches) to any part of the
Aspect
100-4T G3 or
Aspect
ISR G3, including
cables connected to the table. Otherwise, degradation of the performance of this
equipment could result.
NOTE: The emissions characteristics of this equipment make it suitable for use in
industrial areas and hospitals (CISPR 11 class A). If it is used in a residential environment (for
which CISPR 11 class B is normally required) this equipment might not offer adequate protection
to radio-frequency communication services. The user might need to take mitigation measures,
such as relocating or re-orienting the equipment.
If this equipment receives excessive electromagnetic interference, motion controls for
the table may be slow or unresponsive to user inputs. In this event,
1.
Verify the cause by turning nearby equipment off and retest motions. (Note: All
motions will be slowed when operating on battery backup as compared to full
AC power)
2.
If this problem is not resolved, immediately remove power to the equipment by
engaging the Emergency Stop Pushbutton as shown in section 6 of this manual.
3.
Notify IDI customer service using the contact information in the section 6 of
this manual.
CAUTION
Applying pressure to the side of the tabletop more than
70 pounds of force can overcome the braking system
causing the tabletop to move.