Operator Manual
Aspect 100-4T G3 and ISR
G3
L100-2843 28-MAR-2018
31
Guidance and Manufacturer’s Declaration Emissions, Equipment and Systems that
are NOT Life-Supporting
Table 3
The
Aspect
100-4T G3
and the
ISR
G3
are intended for use in the electromagnetic
environment specified below.
The customer or user of the
Aspect
100-4T G3
or the
ISR
G3
should ensure that it is used
in such an environment.
Immunity Test
EN/IEC 60601
Test Level
Compliance
Level
Electromagnetic
Environment
Guidance
Conducted RF
EN/IEC 61000-4-6
Radiated RF
EN/IEC 61000-4-3
AC Mains:
3V, 0.15MHz – 80MHz
6V in ISM band between
0.15MHz and 80MHz
80%AM at 1KHz
[see table 5 of IEC
60601-1-2]
SIP/SOPS:
3V, 0.15MHz – 80MHz
6V in ISM band between
0.15MHz and 80MHz
80%AM at 1kHz
[see table 5 of IEC
60601-1-2]
3V/m
80MHz to 2.7GHz
80%AM at 1kHz
AC Mains:
3V, 0.15MHz – 80MHz
6V in ISM band
between
0.15MHz and 80MHz
80%AM at 1KHz
[see table 5 of IEC
60601-1-2]
SIP/SOPS:
3V, 0.15MHz – 80MHz
6V in ISM band
between
0.15MHz and 80MHz
80%AM at 1kHz
[see table 5 of IEC
60601-1-2]
3V/m
80% AM
Portable and mobile
communications equipment
should be separated from the
Aspect
100-4T G3 or the
Aspect
ISR
G3 by no less than the distances
calculated/listed below:
D=(3.5/3)(Sqrt P)
D=(3.5/3)(Sqrt P)
80 to 800 MHz
D=(7/3)(Sqrt P)
800 MHz to 2.7 GHz
where P is the max power in watts
and D is the recommended
separation distance in meters.
Field strengths from fixed
transmitters, as determined by an
electromagnetic site survey,
should be less than the compliance
levels (V1 and E1).
Interference may occur in the
vicinity of equipment containing a
transmitter.