TROUBLESHOOTING

 

Note: The motion of the Tabletop is fully controlled by user interface with switches, 
buttons and a hand actuated brake release.  In the event of a loss of these motions it is 
expected that the Table Top will remain stationary without any unwanted movement.  

Problem/Symptom 

Possible Cause 

Remedy 

1.  Table controls 
are not functioning.  

1. No Power. 
2. Battery is depleted. 
3. Emergency Stop 
Pushbutton engaged. 
4. Table Top Motion 
Control Panhandle 
Switches disabled. 
5. Control connection. 
6. Electromagnetic 
Interference. 
7. Actuators uncalibrated. 

1. Check electrical outlet. 
2. Connect to AC Power. 
3. Reset Emergency Stop Pushbutton.  
4. Reset Panhandle Switch Lockout 
Pushbutton. 
5. Service Controller connections 
according to Service Manual. 
6. Refer to Section 5 of this manual. 
7. Reinitialize Actuator Controller with 
Handset Control. 

2. Lateral and 
Longitudinal Brakes 
stuttering or 
sticking. 

1. Table has been dormant 
for extended length of 
time. 
2. Contamination on brake 
surfaces, or improper 
cleaning or maintenance 
performed. 

1. Release brakes using Table Top 
Motion Control Panhandle and move 
tabletop in all directions. Working 
tabletop motions may free dragging. 
2. Clean brake surfaces according to 
Service Manual. 

3. Lateral and 
Longitudinal Brakes 
not releasing.  

1. No Power. 
2. Emergency Stop 
Pushbutton engaged. 
3. Battery is depleted. 

1. Check electrical outlet. 
2. Reset Emergency Stop Pushbutton. 
3. Connect to AC Power Outlet. 

4. Longitudinal 
Brakes not 
releasing.  

1. Emergency Stop 
Pushbutton engaged. 
2. Table Top Motion 
Control Panhandle not in 
level Trendelenburg 
position. 
3. Table Top does not stop 
at Trendelenburg Level 
position. 

1. Reset Emergency Stop Pushbutton. 
2. Move Table Top to level 
Trendelenburg position. 
3. Service Trendelenburg Rocker 
Switch according to Service Manual. 

 
 

 

Summary of Contents for Aspect ISR G3

Page 1: ...Made in the USA idi Image Diagnostics Inc Manufacturers of Quality Medical Products Aspect 100 4T G3 and ISR G3 Mobile Image Tables Operator Manual...

Page 2: ...regarding specific requirements and regulations applicable to the use of this equipment Image Diagnostics Inc certifies only the equipment After sale operating practices and safety are the responsibi...

Page 3: ...TTINGS 9 8 2 POWER INDICATOR LIGHT 11 8 3 TABLE MOTION CONTROLS 11 8 4 TABLE TOP MOTION HANDSET CONTROLS 13 8 5 FOOT ACTUATED MOTION CONTROLS 15 9 STANDARD ACCESSORIES 16 9 1 Patient Mattress Pad for...

Page 4: ...MAR 2018 www imagediagnostics com iii 12 MAINTENANCE SERVICE REPAIR 22 12 1 RECOMMENDED PERIODIC PERFORMANCE CHECKS 22 12 2 SERVICE REPAIR STATEMENT 22 13 TROUBLESHOOTING 23 14 DISPOSAL OF COMPONENTS...

Page 5: ...ed could result in death personal injury or damage to the equipment Emergency Stop Pushbutton Electric Shock hazard Information or instructions shown near this symbol must be adhered to in order to pr...

Page 6: ...sentative Model of Table Serial Number of Table Indication of European Conformity for sale in the European Economic Area EEA Item complying to Type B applied part per IEC 60601 1 Date of manufacture o...

Page 7: ...STRUCTIONS All persons using this equipment must fully understand its operation instructions emergency procedures capabilities including total range of motion and be aware of all potential safety haza...

Page 8: ...ipment has not been tested for use with high frequency surgical equipment cardiac defibrillators or cardiac defibrillator monitors Use with such equipment may cause patient burns explosion hazards or...

Page 9: ...essary for the safety of patient and operator CAUTION During operation of the table if any unusual sounds and or erratic movement is observed immediately discontinue use of the table CAUTION Do not us...

Page 10: ...ntennas should be used no closer than 30 cm 12 inches to any part of the Aspect 100 4T G3 or Aspect ISR G3 including cables connected to the table Otherwise degradation of the performance of this equi...

Page 11: ...e familiar with the controls before using them External electrical power connection and disconnection is through the AC power cord and outlet The table will operate on 110 V 230 V or on internal batte...

Page 12: ...g systems utilize a passive magnetic brake it remains engaged energized in the event of a complete loss of electrical power Power is necessary to release the brakes As the braking system utilizes perm...

Page 13: ...gage the break and then transfer the patient onto the Table 7 4 Preparation for Performing CPR Return table to level position retract tabletop to minimize overhang and lower table position to a comfor...

Page 14: ...the STEER position is selected they will pivot as the Table is moved until they are pointing front to back and then they stay locked in that position The Table is in FLOAT mode when the Caster Lock P...

Page 15: ...d unlit when there is no electrical power to the Table or when the Emergency Stop Pushbutton is engaged 8 3 TABLE MOTION CONTROLS All Aspect ISR G3 Aspect 100 4T G3 tables are equipped with tabletop m...

Page 16: ...begin Continually pressing down on the Rocker Switch to the right of this symbol will roll the Table Top in a clockwise direction if looking down the length of the Table from the pedestal end When the...

Page 17: ...R G3 only motions to prevent accidental activation The Table Top Motion Control Unit clamps onto either the left or right Accessory Rail on the Table Top and can be moved along the length of the rails...

Page 18: ...endelenburg tilt function buttons are used the Table Top motion will stop at level position for 1 second before motion begins again When the Table s Trendelenburg tilt position is already level and th...

Page 19: ...l take 1 second before the motion of the Table Top begins ISR G3 FOOT SWITCH CONTROL Foot Switch activated control motions for the Aspect ISR G3 table are ISO Roll Trendelenburg down up and elevate lo...

Page 20: ...Patient Restraint Straps can be used in two different ways depending on where the strap is to be located along the length of the Table Top If the strap is to be used on the end of the table that has...

Page 21: ...side of table 6 length European version with 10mm x 25mm metric rail is available 10 2 Vascular Access Arm Board VAB One piece extra wide carbon fiber arm board for vascular access procedures Mounts t...

Page 22: ...t straps European version for 10mm x 25mm metric rail size is available 10 4 Arm Board Shoulder Mount Carbon fiber base Mounts to table by sliding under Patient Mattress Pad Includes padded arm board...

Page 23: ...by manually pivoting the extension up or down to angle the patient s head To remove the Articulating Headrest Extension from the Tabletop depress the release levers located on the underside of the ex...

Page 24: ...essory rail and the tabletop by clamp 10 9 Anesthesia Screen Holder May require a pair of side rail clamps Flexible Hoop If this is to be used in an area of the tabletop that has no side rails then a...

Page 25: ...on d Power cord Handset Control and Foot Control cords must be plugged in at the table base to protect the inside of the connectors from debris e Remove all pads and accessories f Wipe off any excess...

Page 26: ...which is not user serviceable Replacement must be performed only by a qualified service technician Weekly Check battery operation by disconnecting the AC power and running the tabletop up and down Run...

Page 27: ...6 Refer to Section 5 of this manual 7 Reinitialize Actuator Controller with Handset Control 2 Lateral and Longitudinal Brakes stuttering or sticking 1 Table has been dormant for extended length of ti...

Page 28: ...handle Failure 6 Control connection 7 Actuator s uncalibrated 8 Actuator Failure 1 Check electrical outlet 2 Connect to AC Power 3 Reset Emergency Stop Pushbutton 4 Reset Switch Lockout Pushbutton 5 T...

Page 29: ...le to test response with optional Handset Control or Foot Switch Control if available 9 No response to Handset Control or Foot Switch Control Optional 1 No Power 2 Battery is depleted 3 Emergency Stop...

Page 30: ...sal for recycling The table below lists components typically found in IDI products but varies with model and options COMPONENT ITEM RECYCLING GROUP Actuators Spindle and Motor Housing Cable Metal Stee...

Page 31: ...ISR G3 ONLY Approx 12 with automatic stop at level position Tabletop height range ISR G3 81 3 cm to 111 3 cm 32 in to 44 in without pad 100 4T G3 76 2 cm to 109 2 cm 30 in to 43 in without pad Longitu...

Page 32: ...0Hz or 230 5 Vac 50Hz Duty Cycle 10 2 min on 18 min off Current Rating Less than 10 Amps Battery Backup Power Environmental Operating Temperature Range 10 C to 40 C Operating Humidity Range 30 to 75 r...

Page 33: ...G3 should ensure that it is used in such an environment Emissions Test Compliance Electromagnetic Environment Guidance RF Emissions CISPR 11 Group 1 The Aspect 100 4T G3 and the ISR G3 use RF energy o...

Page 34: ...cles for 50 Hz and 60Hz respectively Single phase at 0 0 UT 250 300 cycle for 50 Hz and 60 Hz respectively Single phase at 0 0 UT 0 5 cycle At 0 45 90 135 180 225 270 and 315 0 UT 1 cycle 70 UT 25 30...

Page 35: ...80MHz 6V in ISM band between 0 15MHz and 80MHz 80 AM at 1kHz see table 5 of IEC 60601 1 2 3V m 80MHz to 2 7GHz 80 AM at 1kHz AC Mains 3V 0 15MHz 80MHz 6V in ISM band between 0 15MHz and 80MHz 80 AM a...

Page 36: ...output power of the Communications Equipment Table 4 Max Output Power Watts Separation m 150kHz to 80MHz D 3 5 3 Sqrt P Separation m 80 to 800MHz D 3 5 3 Sqrt P Separation m 800MHz to 2 5GHz D 7 3 Sq...

Reviews: