Operator Manual
Aspect 100-4T G3 and ISR
G3
L100-2843 28-MAR-2018
28
16.
SPECIFICATIONS
Mode of Operation
▪
For continuous use with short time loading.
▪
Duty Cycle: 10% (2 min on/18 min off).
Type of Equipment
:
▪
Class II Type B applied part (as defined by IEC 60601-1, UL 60601-1,
EN 60601-1, CAN/CSA 601.1-M90, IEC 60601-2-46:1998.
▪
Type B protection against electrical shock as the applied part is the table surface.
Electrical:
▪
Supply Voltage: 120±5% Vac 60Hz or 230
5% Vac 50Hz.
▪
Duty Cycle: 10% (2 min on/18 min off).
▪
Current Rating: Less than 10 Amps.
▪
Battery Backup Power.
Environmental
:
▪
Operating Temperature Range: -10
C to +40
C.
▪
Operating Humidity Range: 30% to 75% relative humidity, noncondensing.
▪
Operating Pressure Range: 700 hPa to 1060 hPa.
▪
Transport & Storage Temperature Range: -40
C to +60
C.
▪
Transport & Storage Humidity Range: 30% to 75% relative humidity, noncondensing.
▪
Transport & Storage Pressure Range: 500 hPa to 1060 hPa.
▪
Rated IPX4 (Protected against splashing water).
▪
Meets EMC requirements of IEC 60601-1-2:2007.
Tabletop:
▪
The tabletop is made of carbon fiber and meets all the requirements of FDA CFR Title
21, Chapter 1, Subchapter J.