Operator Manual
Aspect 100-4T G3 and ISR
G3
L100-2843 28-MAR-2018
25
Problem/Symptom
Possible Cause
Remedy
8. No response to
Table Top Motion
Control Panhandle
controls.
1. No Power.
2. Battery is depleted.
3. Emergency Stop
Pushbutton engaged.
4. Table Top Motion
Control Panhandle
Switches disabled.
5. Table Top Motion
Control Panhandle cable
damaged or Panhandle
failure.
1. Check electrical outlet.
2. Connect to AC Power.
3. Reset Emergency Stop Pushbutton.
4. Reset Switch Lockout Pushbutton.
5. Replace Panhandle Unit.
(It is possible to test response with
optional Handset Control or Foot
Switch Control if available.)
9. No response to
Handset Control or
Foot Switch Control
(Optional).
1. No Power.
2. Battery is depleted.
3. Emergency Stop
Pushbutton engaged.
4. Handset Control or
Foot Switch Control
Cable damaged.
5. Actuator(s)
uncalibrated.
6. Device Failure.
1. Check electrical outlet.
2. Connect to AC Power.
3. Reset Emergency Stop Pushbutton.
4. Test response with Table top Motion
Control Panhandle controls.
5. Reinitialize Actuator Controller with
Handset Control.
6. Replace faulty device.
10. Elevate,
Trendelenburg and
ISO-Roll slow or
unexpected
movement.
1. Running on Battery
Power.
2. Actuator(s)
uncalibrated.
1. Connect to AC Power.
2. Reinitialize Actuator Controller with
Handset Control.