Installation and Maintenance Guide for IDI Suspension System mounted to Canon DST-1000A Rails
L240-1429 REV A Image Diagnostics, Inc.
Page
22
YOKE BALANCE ADJUSTMENT
WARNING!
Yoke has been leveled at manufacturers. Make
adjustment only if absolutely necessary to compensate
for an unbalanced load.
NOTE: Prior to yoke adjustment,
payload must be secure and arm
horizontal.
To compensate for yoke balance, fine adjustments can be made by turning the link. Loosen
threaded hex nut at outboard rod end, spin link until yoke is in a horizontal position and then
retighten hex nut. See Figure 17.
Figure 15
MONITOR MOUNT ASSEMBLY ROTATION RESISTANCE
1.
The setscrews shown in Figure 6 that were installed in step 6 of "INSTALL MONITOR
MOUNT" on page 12 can be loosened or tightened to adjust the rotation resistance of the
suspended monitor mount assembly.
2.
Replace wireway cover after making adjustment.
F. CLEANING THE EQUIPMENT
No part of this unit is designed to be sterilized in an autoclave. Do not allow water or other
liquids to permeate inside the equipment as this may cause short circuits or corrosion. Clean
parts with a clean cloth dampened with disinfectant or a mild detergent solution. Do not use
abrasives, solvents, sprays or corrosive cleaning agents. Gently rub with a clean soft cloth
to dry.
If room is to be disinfected by means of an atomizer, the equipment must be covered with
plastic or similar sheeting. The equipment must be turned off well in advance of this
procedure to prevent convection currents from drawing the disinfectant mist into the
equipment. After the mist disperses completely, the sheeting may be removed, and the
equipment disinfected as described above.