idi 1000F-1 Installation And Maintenance Manual Download Page 19

Installation and Maintenance Guide for IDI Suspension System mounted to Canon DST-1000A Rails 

L240-1429 REV A                                        Image Diagnostics, Inc. 

 

                                        Page 

19 

INSTALL MOBILE CABLE HANDLING ASSEMBLY 

 
1.

 

Slide two Nut Plates into the outer track of the Cross Rails.  Apply Loctite #242 to the threads 
of two standoffs and then assemble into nut plates hand tight only. See Figure 12. 

 

Figure 10

 

2.

 

Align Safety Brackets with the inner threaded holes of the Nut Plates and fasten with a lock 
washer and a 3/8in. long socket head cap screw.  

3.

 

Slide Cable Trolley Assembly onto Cable Hanger shaft.  See Figure 13. 

4.

 

Assemble Cable Hanger shaft through holes in Safety Brackets and into the side holes of each 
shaft support. 

5.

 

Install  mounting  screw  through  the shaft supports,  through the hole in  the end of the  Cable 
Hanger shafts and into the standoffs.  Lightly tighten mounting screws. 

 

Figure 11 

 

 

Summary of Contents for 1000F-1

Page 1: ...Suspension System Model IDI 1000F 1 and 1000F 2 with Adapter for Canon DST 1000A Ceiling Rails Installation and Maintenance Guide The text of this manual was originally written approved and published...

Page 2: ...Y STOP 10 INSTALL CROSS RAIL END PLATE 11 END LOAD MONITOR MOUNT CARRIAGE 12 INSTALL MONITOR MOUNT 13 INSTALL MONITOR S 14 SET ROTATIONAL LIMITS 16 ELECTRICAL CONNECTIONS 17 INSTALL STATIONARY CABLE H...

Page 3: ...dent panel mounting and positioning This mobile package includes clip mounted bridge rails PRODUCT DATA Package Includes o Mobile gantry and counterbalanced swing arm with 42 sweep radius 25 vertical...

Page 4: ...t in death personal injury or damage to the equipment Electric Shock hazard Information or instructions shown near this symbol must be adhered to in order to prevent a potentially hazardous situation...

Page 5: ...Canon DST 1000A Rails L240 1429 REV A Image Diagnostics Inc Page 5 Indication of European Conformity for sale in the European Economic Area EEA Item complying to Type B applied part per IEC 60601 1 D...

Page 6: ...eral and local recommendations concerning electrical safety when used in medical installations All electrical connections shall be done by licensed approved electrician per national electric codes CAU...

Page 7: ...ge Diagnostics Inc 310 AUTHORITY DRIVE FITCHBURG MA 01420 USA Or call IDI at 978 829 0009 or send fax to 978 829 0027 Or call Toll Free at 877 304 5434 SAFETY HAZARDS ALERTS Alert Circumstances for us...

Page 8: ...ations INSTALL DST 1000A CEILING RAIL ADAPTERS Before Installing the Canon supplied longitudinal ceiling rails make sure that there is at least 35 in 89 cm between one end of the rails and any obstruc...

Page 9: ...s to each of the two rail adapters in the inner rail clip locations as shown in Figure 2 using 3 8 16 x 2 painted bolts flat washers lock washers and nylon locknuts Make sure to orientate the Rail Cli...

Page 10: ...ip hardware using Loctite 242 or equivalent INSTALL RAIL X Y STOP 1 The X Y stop unit should have a T nut fastened to it with two socket head cap screws Loosen the screws for the T nut and insert the...

Page 11: ...STALL CROSS RAIL END PLATE 1 Install only one Cross Rail End Plate sub assembly at the far end of the Cross rails making sure that the rubber bumpers face in towards the rails See Figure 5 Use Loctite...

Page 12: ...re all carriage wheels are engaged onto rails before lowering Hi Jack 3 Remove pallet and shipping bars from support unit 4 If a Stationary Mobile Cable Handling Assembly is going to be used with this...

Page 13: ...aring from inside the monitor mount frame 4 Install the thinner stainless steel thrust ring onto the rotation bearing from inside the monitor mount frame 5 Install six socket head cap screws with lock...

Page 14: ...ach monitor to be installed insert screws into the top two mounting holes leaving a gap below the head of the screw big enough to fit over the thickness of the mounting bracket to which it is being mo...

Page 15: ...ce by hand tighten the socket head cap screw on the pivot point of the bracket NOTE TO INSTALLER Ground studs have been provided to properly ground monitor s Terminal strips have been provided for dis...

Page 16: ...nted to Canon DST 1000A Rails L240 1429 REV A Image Diagnostics Inc Page 16 SET ROTATIONAL LIMITS If desired the rotational stop pin bumper can be relocated to one of four possible mounting locations...

Page 17: ...nylon cable ties Tying the cord into a knot or tying the ends with string shall not be used for cord anchorage 3 Arrange the conductors of the power supply cord so that the protective earth conductor...

Page 18: ...rail track prior to positioning the Mobile Cable Handling Assembly If assembly is to be positioned at the near end of rail slide into the rail after positioning the Mobile Cable Handling Assembly The...

Page 19: ...nd tight only See Figure 12 Figure 10 2 Align Safety Brackets with the inner threaded holes of the Nut Plates and fasten with a lock washer and a 3 8in long socket head cap screw 3 Slide Cable Trolley...

Page 20: ...adjust the exposed length of the setscrew on the standoff DRESS CABLES 1 When the monitors have been properly installed it is recommended that the cables run through the pivot tube as shown in Figure...

Page 21: ...ial settings from gas spring data label See Figure 15 Figure 13 2 Prior to adjustment payload must be secure The arm can be pulled down to make access to adjustment nut easier Fine adjustment to compe...

Page 22: ...can be loosened or tightened to adjust the rotation resistance of the suspended monitor mount assembly 2 Replace wireway cover after making adjustment F CLEANING THE EQUIPMENT No part of this unit is...

Page 23: ...onal Checks Perform daily checks of the monitor suspension BEFORE and AFTER operating the equipment CAUTION If any abnormality is found in the monitor suspension stop using it Post a sign reading DO N...

Page 24: ...no abnormal sounds The monitors or monitor suspension are not tilted due to fasteners that are not tightened properly Annual Maintenance Check 1 Check the condition of the wheels on the adapter and c...

Page 25: ...s 5 Clean and touch up the painted surfaces 6 Refer to Figure 22 for the following instructions Check that the hardware is tight and that there is no wear at A the end of the gantry core arm B the rot...

Page 26: ...it has been repaired Operating equipment with defective components may expose the operator or the patient to safety hazards The manufacturer strongly recommends that a maintenance program be initiate...

Page 27: ...of metal usually aluminum or steel and are easily recycled The gas springs are nearly 100 recyclable They must be decompressed with the remaining oil drained before processing The compressed gas insi...