36
5.6.2 Max/min
Max/min is the section of the menu to clear the maximum and minimum values in the main display
screen. These values will begin at when the analyzer is returned to the display screen.
To clear the maximum and minimum values:
1) Use the up
[
5
]
and down
[
6
]
arrow buttons to highlight
MAX/MIN
and press the enter
[ENTER]
button. The analyzer will ask for a confirmation before proceeding.
2) Use the left
[
3
]
and right
[
4
]
arrow buttons to toggle between
YES
and
NO
and press enter.
3) Choosing no will return the analyzer to the clear menu without clearing the max and min values.
4) Choosing yes will clear the max and min values and return the analyzer to the clear menu.
5.7 Off
menu
From the off menu, the user can enable the password to protect the power analyzer from accidental shut
off. By activating the turn-off password protection, the user will be prompted for a password before the
analyzer can be shut down.
To change the password protection status:
1) Use the up
[
5
]
and down
[
6
]
arrow buttons to move the cursor through the available parameters of
the programming menu, select
OFF
and press the enter
[ENTER]
button.
2) The analyzer will prompt the user to activate the password protection
3) Use the left
[
3
]
and right
[
4
]
arrow buttons to toggle between
YES
and
NO
and press enter.
5) Choosing no will deactivate the turn-off password protection and return the unit to the main
programming menu.
6) Choosing yes will activate the turn-off password protection and return the unit to the main
programming menu.
6. Hooking up the analyzer
Before hooking up the 800 series power analyzer, please observe the following:
•
The unit has sufficient battery life for the duration of the test, or is connected to line power.
•
Supply voltage is connected to a properly grounded outlet.
•
Maximum input voltage does not exceed analyzer ratings:
o
500VAC phase to neutral
o
866VAC phase-phase
•
Clamp adapter range sufficient for the measurement circuit
o
Does not exceed maximum current rating of adapter
o
Does not fall below 10% of current rating of adapter
•
Voltage leads being used were supplied with the instrument. Replacement leads are available from
IDEAL INDUSTRIES
•
Power analyzer has been setup for the electrical system under test