5. Setting up the device
12
Menu level
Parameterisation level
Impedance matching,
:
Default:
The value for the sensor calibration is aligned from the smallest to the highest digit
[▼] [▲]
and confirmed digit per digit with
[P]
. After the last digit, the device
changes back into menu level. During a temperature measurement in
°
C the value
calibration can be adjusted between -20.0 and +20.0 and can be set during a
measurement in
°
F between -36.0 and +36.0. If the measurement is switched later
on, the value is rounded.
Setting up the display time,
:
Default:
The display time is set with
[▲] [▼]
. The display moves up in increments of 0.1
sec up to 1 sec and in increments of 1.0 sec up to 10.0 sec. Confirm the selection
by pressing the
[P]
button. The display then switches back to the menu level
again.
Device undercut,
Default:
With this function the device undercut (
) can be defined on a definite value.
Exception is input type
4-20 mA
, it already shows undercut at a signal <1 mA, so a
sensor failure is marked.
Display overflow,
Default:
With this function the display overflow (
) can be defined on a definite value.
Back to menu group level,
With
[P]
the selection is confirmed and the device changes into menu group level
.
then