7
5. Setting up the device
Menu level
Parameterisation level
Impedance matching,
:
Default:
The value for the sensor calibration is aligned from the smallest to the highest digit
[▼] [▲]
and confirmed digit per digit with
[P]
. After the last digit, the device
changes back into menu level. During a temperature measurement in
°
C the value
calibration can be adjusted between -20.0 and +20.0 and can be set during a
measurement in
°
F between -36.0 and +36.0. If the measurement is switched later
on, the value is rounded.
Setting up the display time,
:
Default:
The display time is set with
[▲] [▼]
. The display moves up in increments of 0.1
sec up to 1 sec and in increments of 1.0 sec up to 10.0 sec. Confirm the selection
by pressing the
[P]
button. The display then switches back to the menu level
again.
Selection of analog output,
Default:
Three output signals are available: 0-10 VDC, 0-20 mA and 4-20 mA, with this
function, the demanded signal is selected.
Setting up the final value of the analog output,
Default:
The final value is adjusted from the smallest digit to the highest digit with
[▲] [▼]
and digit by digit confirmed with
[P]
. A minus sign can only be parameterised on
the highest digit. After the last digit, the device changes back into menu level.
then