5-5
5
TABLE 5-2 TROUBLESHOOTING GUIDE (CONT.)
Possible
Symptom
Fault
Action or Check
8099 has to be reset
Program uses all
Check program for excess
often.
8099 resources
opening and closing links.
Rewrite program to keep links
open until program ends.
COMM_Timeout
8099 not accessed for a long
expired
time period. Extend 8099
COMM_Timeout period or
add a background keepalive
functon to the client program.
.
8099 stops
Using Agilent IO
Agilent expects their
responding
Libraries
instruments to disconnect
when the link count goes to
zero. Turn Auto-disconnect
on for operation with this kind
of program.
Note that opening and closing
sockets and links is a time
wasting operation and slows
down your program. If
possible, rewrite the program
to avoid repeatedly opening
and closing sockets and links.