![ICS 8099 Instruction Manual Download Page 46](http://html1.mh-extra.com/html/ics/8099/8099_instruction-manual_3540995046.webp)
3-10
3
TABLE 3-2 IEEE-488.2 COMMON COMMANDS
(CONTINUED)
COMMAND NAME
DESCRIPTION
*IDN?
*OPC
*OPC?
*PSC<value>
*PSC?
*RCL
<value>
Identification
Query
Operation
Complete
Command
Operation
Complete
Query
Power-On
Status Clear
Power-On
Status Clear
Query
Recall
8099 returns its identification code as
four fields separated by commas. These
fields are: manufacturer, model, six-digit
serial number and hardware-firmware ver-
sion and date e.g.
ICS Electronics, 8099,
S/N 710001, Rev. 00.X0, 07.10.24.
The
IEEE-488.2 specification states that the
word 'model' may not appear in the IDN
message.
Causes the 8099 to generate the operation
complete message in the Standard Event
Status Register when all pending selected
8099 operations have been finished.
Places an ASCII character 1 into the 8099's
Output Queue when all pending selected
8099 operations have been finished.
Controls the automatic power-on clearing
of the SRE and ESE registers. *PSC 0 al-
lows devices to restore the saved SRE and
ESE values and to assert SRQ upon power
turn-on. *PSC 1 enables the power-on clear
and disallows a SRQ at power turn-on. The
PSC commands saves the 488.2 SRE and
ESE registers and the SCPI transition and
enable register values.
Querys the PSC flag value. A returned value
of 0 indicates the registers will retain their
saved values, a returned value of 1 indicates
the registers will be cleared.
Restores the state of 8099 from a copy
stored in its Flash by *SAV command.
*RCL 0 recalls saved configuration, up-
dates output levels and re-initializes the
UART. Allow the 8099 300 ms to complete
the *RCL command.