![ICS 8099 Instruction Manual Download Page 52](http://html1.mh-extra.com/html/ics/8099/8099_instruction-manual_3540995052.webp)
3-16
3
:CONDition?
Returns contents of the condition regis-
ter
associated with the command.
:ENABle
0
Sets the enable mask which allows the true
conditions in the associated event register
to be reported in the summary bit.
:PTRansition
#h7FFF
Sets positive transition enable regis-
ter. Value = 0 to #h7FFF in decimal or
HEX.
:NTRansition
0
Sets the negative Transition register.
Values = 0 to #h7FFF in decimal or
HEX.
:PREset
Sets the selected Enable Register, PTR
and NTR registers to their default values
(0, #h7FFF)
FORMat
Starts string format branch.
:DATA
Optional digital data identifier
:TALK
ASCii
Sets talk string and data query response
format. ASCII expresses a words input
bit pattern as a decimal value equal to the
binary sum of the data. Multiple words
are separated by commas. HEXL converts
each four bit nibble into the ASCII char-
acters 0-9 and A-F. TABLE allows the
user to define his own character set. All
talk strings end with a linefeed. Values
are ASCii | HEXL |.
i.e. ASCii example = 128,5,255
HEXL example = 8000, 05FF
TABLE 3-3 SCPI COMMANDS AND QUERIES
(CONTINUED)
Keyword
Default
Description
Value