11
5. Setting up the device
Menu level
Parameterisation level
Setting the lower value of the analog output,
Out.OF:
Default:
00000
The final value is adjusted from the smallest digit to the highest digit with
[
▲
] [
▼
]
and digit by
digit confirmed with
[P]
. A minus sign can only be parameterised on the highest digit. After the
last digit, the device changes back into menu level.
Threshold values / limit values,
LI-1:
Default:
2000
For both limit values, two different values can be parameterized. With this, the parameters for
each limit value are called up one after another.
Hysteresis for limit values,
HY-1:
Default:
00000
For all limit values exists a hysteresis function, that reacts according to the settings (threshold
exceedance / threshold undercut).
Function for threshold value exceedance/undercut,
FU-1:
Default:
high
The limit value undercut can be selected with
Louu
(LOW = lower limit value) and limit value
exceedance can be selected with
high
(HIGH = upper limit value). If e.g. limit value 1 is on a
switching threshold of 100 and occupied with function
high
, the alarm will be activated by
reaching the threshold. If the limit value is allocated to
Low
, an alarm will be activated by
undercut of the threshold. See page 29.
Threshold values / limit values,
LI-2:
Default:
3000
This value defines the threshold, that activates/deactivates an alarm.
Hysteresis for limit values,
HY-2:
Default:
00000
The delayed reaction of the alarm is the difference to the threshold value, which is defined by
the hysteresis.