Instruction Manual C200
3.8.2 System errors
System errors are displayed in the event of hardware problems. These error messages assist
our service department in analysing the problem. Switch off the device and then switch it on
again. If the error message recurs, please contact the
IBR service department
.
Error
Error description
Corrective action
no.dAtA
The C200 does not receive any measure-
ment data. Possible causes :
1. No IMBus modules have been connected,
or an IMBus module is defective.
2. Cascading is not possible because the
C200 with the IMBus measuring and
interface modules is still turned off.
1. Check IMBus modules.
2. Turn on all C200s for cascading.
Err. 97
The bootloader version of a connected
IMBus module is < 1.2.
No correction possible. The connected IMBus
module is not compatible with the C200.
Err. 98
The firmware version of a connected
IMBus module is < 1.6.
Update IMBus module firmware.
Err. 99
Settings were not saved correctly.
Repeat previous action.