29
Instruction Manual C200
3.6 Basic setup
The Basic Setup menu comprises all basic device settings which are usually only programmed during the
initial operation.
The basic factory setting is identified by an
∗
in the individual menus
.
GAGES
Push the encoder button to select the menu. Turn the encoder to browse through the individ-
ual gauges from
C1
to
C8
. The numeric display will indicate the number of the gauge and its
current status
(
on / oFF
).
To change the status of a gauge, push the encoder button and then turn the encoder to select
the new status. Push the encoder button again to confirm the new status.
Note :
If all gauges have been switched off (
oFF
), the gauge
C1
is automatically
switched on again when exiting the programming menu, because measurement
operation is not possible without an active gauge.
Auto.rE.
In this menu you can choose either
Manual gauge selection
or
Automatic gauge selection
by auto recognition
.
Push the encoder button to select the menu. The numeric display indicates
Aut.oFF / on
.
Turn the encoder to activate (
on
) or deactivate (
oFF
) the automatic gauge selection
function. Push the encoder button again to confirm the new status.
Note :
If one measuring input
P1
to
P8
is assigned to several gauges in the operating
mode
Gauge selection by auto recognition
, the numeric display of the column
gauge will indicate
‘Err. 08’
.
If a measurement value would change for a measuring input that is assigned several
times, the measurement values for several gauges would change at the same time,
and an automatic gauge selection would consequently not be possible.