26
Instruction Manual C200
The
MEASURING INPUTS
menu enables assigning the measuring inputs (
P1
to
P8
) to the currently se-
lected gauge (
C1
…
C8
).
The 8 inputs can be linked in any order
( e.g.: P1+P2, P1-P2,
P1+P2 - P3+P4, etc. ).
Every measuring input can be multiplied by a “device coefficient” ( multiplier ) ranging from 0.001 to
59.999. Enter
1
as the multiplying factor, if the input is to be added without correction; enter
-1
, if the
input is to be subtracted without correction.
Examples :
a)
Bore gauge :
Ø d1 = - P1
P1
-
1.000
P2
off
P3
off
P4
off
P5
off
P6
off
P7
off
P8
off
b)
Shaft measurement :
Ø d1 = P1 + P2
P1
1.000
P2
1.000
P3
off
P4
off
P5
off
P6
off
P7
off
P8
off
c)
Taper measurement :
Taper = Ø d2 – Ø d1
Taper = ( P3 + P4 ) - (
P1 + P2 )
Taper =
P3 + P4 - P1 - P2
P1
-1.000
P2
-1.000
P3 1.000
P4 1.000
P5
off
P6
off
P7
off
P8
off