IBR C100 Instruction Manual Download Page 32

 

 
                                                                                                                                                                                
                                                                                                                                                                  

 
 

                                                       

     32      

 

Instruction Manual     C100 

    7.    Accessories and order information 

   

   Designation

                                                                                      

Article-Number

 

   

 

C100 

Column gauge without

 

measuring module, 

Dynamic mode

  

F201 

001 

  

 

 

Multi gauging mode 

 

F201 002 

 

 

Grading mode

  

F201 

003 

 

 

Flatness mode

  

F201 

004 

 

C100-AE 

Column gauge incl. pneumatic measurement converter, air filter  

 

F202 001 

 

and pressure regulator 

 

 

 

 

 

6-digit numeric display, 100 bar column display ( 3-colour ) 

 

LED´s for unit, range and function display, RS232, trigger input 

 

digital outputs, switching power supply for 

~

85...260V, incl. 

 

footplate and manual. 

 

B/C100-ind2 

B/C100 measuring module to connect 2 inductive probes  ..............................  F220 

010 

 

2 analogue inputs and outputs ( Tesa halfbridge and compatible ) 

 

B/C100 measuring module to connect 2 inductive probes,  .............................  F220 

011 

 

2 analogue inputs and outputs ( each inductive probe types ) 

B/C100-ind4 

B/C100 measuring module to connect 4 inductive probes   .............................  F220 

020

 

( Tesa halfbridge and compatible ) 

 

B/C100 measuring module to connect 4 inductive probes ...............................  F220 

021 

 

( each inductive probe types )

 

B/C100-ana4

 

B/C100 measuring module to connect 4 analogue inputs  ...............................  F220 

030

 
 

AE1 

Pneumatic measuring converter with analogue voltage output

  

F330 

001 

 

AE-FC1 

Connection set for 1...4 AE1 pneumatic measuring converters

   

 

F330 010 

 

Connection cable, coupling for air-connection, air-filter, mounting bracket 

 

Special pressure regulator and T-connectors with 3m air hose 

 

AE1-FP 

Footplate for 1…4 AE1 pneumatic measuring converters                             

F330 020 

 

AE1-Adapt 

Adapter cable to connect several B100 / C100 gauges

   

 

F330 030 

 

on an AE1-bus 

 
CA2 

Gauge to gauge connection adapter with terminal block for 2 outputs and 

 

F220 090 

 

2 inputs with adjustment 

CA4

 

Gauge to gauge connection adapter with terminal block for 4 inputs with

 

  F220 

091 

 adjustment 

 
 

PC-RS232 cable 

Connection cable for RS232 with adapter 9/25 pin ( COM1…COM4 )  ............  

F601 002 

PC-USB cable

 

Connection cable for PC USB with driver software to emulate COM 1…127  ..  

F601 020 

Profibus 

Profibus connection module

 

Unigate IC ( External Gateway ) .........................  F601 

071 

 

Profibus connection module Unigate SC ( External Gateway ).........................  

F601 072 

 

 

RSD

 

Connection cable for Mitutoyo Digimatic inputs ...............................................  

F601 030 

 

( cable change OPTO RS232 to DIGIMATIC ) 

RSA

 

Analogue output adapter with plug-in terminal block .......................................  

F601 031 

 

( adapter change OPTO RS232 

 ±10V voltage output ) 

 
Foot switch 

Foot switch, protection IP32  ............................................................................  F602 001 

 

Foot switch, protection IP65  ............................................................................  F602 002

 

Foot switch with adapter connector, protection IP32 .......................................  

F602 003

 

Foot switch with adapter connector, protection IP65 .......................................  

F602 004

 

 

 
 
 

Summary of Contents for C100

Page 1: ...M e t r o l o g y Document No D2F201 002 Edition June 2003 Copyright IBR Messtechnik GmbH Co KG INSTRUCTION MANUAL C100 Column gauge Multi gauging...

Page 2: ...for programmers in a hurry 12 3 4 Programming menus 13 3 5 Basic settings Setup menu 17 3 6 Restoring factory settings 21 3 7 Changing system settings 21 3 8 Error messages Error corrections 21 4 Work...

Page 3: ...pact design of the microprocessor based electro nics allows manual and automatic selection of 1 4 gauges to measure different work pieces or characteristics in an industrial environment Manual gauge s...

Page 4: ...olerance limits are marked on the column display 0 0050 mm 0 00050 Notice The measurement range of the column display can be set in the Setup menu to a fixed value or to Auto mode In AUTO mode the col...

Page 5: ...5 Instruction Manual C100 102 mm 136 5 mm 390 mm 370 mm 45 mm 52 mm 5 1 4 Dimensions 1 5 Connectors round up all accessories shown Figure Column display on footplate...

Page 6: ...tic or analogue sensors 2 DIN45322 sockets 5 pin and one 9 pin Sub D socket with 2 analogue inputs and 2 outputs 1 5V 10000 Digit 4 DIN45322 sockets 5 pin 9 pin Sub D socket 4 analogue inputs 1 5V 100...

Page 7: ...re use 2 2 Fitting the footplate Fix the footplate to the instrument and place on a sturdy and level surface 2 3 Connecting inductive probes Compare the probe type with the specifications given on the...

Page 8: ...ch are connected at the rear of the column gauges The potentiometers mounted in the adapters allow an individual adjustment of incoming measurement signals Pin assignments are shown in chapter 6 1 and...

Page 9: ...Setup menu and is identical for both connectors RS232 and Tol Out Foot and hand switch types Art No Foot switch IP32 F602 001 Foot switch IP65 F602 002 Foot switch with adapter connector IP32 F602 00...

Page 10: ...th plug in terminal block F601 031 adapter converts OPTO RS232 10V analogue output 2 9 Connecting to the power supply First make sure that the rated voltage complies with the electrical requirements i...

Page 11: ...to increase a given value or switch to the next menu or function When this key is pressed in Measuring mode the measured value on the numeric display is reduced A short pressure decrements by 1 If th...

Page 12: ...50 m or 0 1500 inch 0 0005 inch c Controlling of display output by foot switch oFF trAnS hold d Automatic zero adjustment using keypad on oFF using foot switch on oFF e Serial data transfer continuou...

Page 13: ...key to advance to the next menu In this menu you can switch on or off the gauges 1 4 and access both Programming mode for the gauges 1 4 and the SetUP menu Press the and keys to select the gauges 1 4...

Page 14: ...s the measured value and appears on the numeric display When measuring comparative values shows the deviation to the nominal size on numeric and column displays zero is programmed as master value When...

Page 15: ...can set a maximum of 4 tolerance points representing relative deviation from the nominal value Programming starts by selecting one of the four tolerance marks on the column display red upper toleranc...

Page 16: ...0 00 D 0 00 D b A B 1 00 A d A 2B C 1 00 A 1 00 B 2 2 00 B 0 00 C 0 50 C 0 00 D 0 00 D Notice When all probe inputs are disabled 0 00 A D measurements are no longer possible and the error message Err...

Page 17: ...n a new resolution is set only the decimal point of programmed values is shifted The programmed numeric values e g master values nominal size and tolerances must be corrected in some cases rAnGE In th...

Page 18: ...are no longer available because the foot switch is needed to start stop dynamic measurements or for gauge switching Error messages Err 3 appears on the numeric display 0 triG Enable or disable Automa...

Page 19: ...Current measured value transmitted by foot switch AutoCA In this menu you can select between the options Automatic zero adjustment using one master and Automatic gauge calibration using two masters on...

Page 20: ...the numeric display will force the operator to input a password consisting of a 6 digit number Notice If the Programming menu is called later by the PRG CAL key the column gauge will prompt you to int...

Page 21: ...5mm or 0 059 inch L rAnG 15mm or 0 59 inch for long travel probes with 1 10 of the sensitivity System Setup mode is terminated by pressing the PRG CAL key 3 8 Error messages Error corrections Operati...

Page 22: ...ts On small master differences use gauge in High Resolution mode Press PRG CAL key to enter Programming mode Choose menu Set Master value s and introduce 2 masters with a difference 6 digits PrESEt in...

Page 23: ...djustment and the Automatic gauge calibration is done in the Setup menu AutoCA 4 3 Mechanical adjustment of inductive probes Inductive probes reach their greatest accuracy only within a comparatively...

Page 24: ...down the measured value rapidly increases or decreases If SECUrE appears in the numeric display the keypad is locked Refer to Setup menu Automatic adjustment 0 triG Adb on oFF set to on 4 5 Automatic...

Page 25: ...s Cal dongel During the calibration the 2 masters are placed alternately in the measuring fixture and the difference between the 2 measured values is calculated Use the relevant potentiometer to adjus...

Page 26: ...ta format The data format has a fixed length of 10 digits including the end character cr ASCII 13 The measured value always starts with a sign followed by leading zeroes The decimal point is shifted a...

Page 27: ...er multiplexer sends a cr to the column gauge to request data Example PC Programme DOS BASIC 100 REM REQUEST MEASUREMENT VALUE BY 110 OPEN COM1 4800 E 7 2 FOR RANDOM AS 1 open COM device 120 PRINT 1 C...

Page 28: ...t values into MS EXCEL The IBREXDLL XLS is an EXCEL workbook for the component or characteristic depending data collection into Excel tables Therefore extensive possibilities for data collection forma...

Page 29: ...29 Instruction Manual C100 6 Pin assignments of male connectors 6 1 Connectors of the column gauge...

Page 30: ...30 Instruction Manual C100 6 2 Plug in terminal blocks for accessories...

Page 31: ...31 Instruction Manual C100...

Page 32: ...output F330 001 AE FC1 Connection set for 1 4 AE1 pneumatic measuring converters F330 010 Connection cable coupling for air connection air filter mounting bracket Special pressure regulator and T conn...

Page 33: ...Accu module incl accu F220 110 for B100 and C100 F220 100 On exchange with integrated power supply module Universal charger for Lithium Ionen Accus VW VBD1E F220 102 Lithium Ionen Accu Serie VW VBD1E...

Page 34: ...or written consent of IBR Messtechnik GmbH Co KG and must be carried out by competent staff Opening the case or tampering with the device without prior permission will lead to the loss of the guarante...

Page 35: ...f delivery This guarantee covers all materials and manufacturing defects Our liability is confined to repair or should we deem it necessary replacing or crediting the goods The following are not cover...

Reviews: