Instruction Manual C100
After pressing the
ENTER
key ( confirm flashing "Unit" ) in the 2nd menu you can select either „
mm
“ or
„
INCH
" by pressing the
Ï
or
Ð
key. Confirm with the
ENTER
key
.
Press the
Ð
key to advance to the next menu.
Notice :
a) If you change the unit, numeric values introduced in the programme, e.g. master values,
nominal size values and tolerances do not change automatically.
b) Further units on request.
The programming of a master value is base of the Automatic Zero Adjustment.
On Automatic zero adjustment
( do this by placing the master piece into the measuring fixture and press
the
Ï
and
Ð
keys simultaneously during Measuring mode
) the master value will be taken as the
measured value and appears on the numeric display.
When measuring comparative values ( shows the deviation to the nominal size on numeric and column
displays ) zero is programmed as master value.
When pneumatic converters are used, Automatic Gauge Calibration can be considered as an option. If
this option is chosen in the Setup menu
"AutoCA"
you will be prompted to program a second master
value.
If you then enter the same value for both masters, the error message
'Err. 11'
will appear. If you enter a
difference between the two masters with more than 3 mm the error message
'Err 12'
will appear.
Notice :
On Auto Zero Adjustment the zero position of the column gauge related to the master is
automatically adjusted by an offset value.
On Automatic gauge calibration the offset value ( zero point ) and the gain ( pneumatic
spread ) is calculated automatically by the column gauge.