11
Instruction Manual C100
3. Programming the column gauge
All programming and settings of the column gauge are done with the 4 keys on the front panel. The user
is taken step by step through the different menus with the help of the lamps and the numeric display. The
programming follows a logic structure and becomes self-explanatory after a short study.
3.1 Key functions
PRG/CAL
When this key is pressed the gauge changes from
Measuring
to
Programming mode
or
vice versa.
ENTER
On
Measuring mode
( manual gauge selection mode ) with the ENTER key you switch
to the next gauge.
In
Programming mode
this key confirms the programmed values and selections.
Ï
When this key is pressed in
Measuring mode
the measured value on the numeric
display is increased. A short pressure increments by 1. If the key is held down the
measured value is increased rapidly ( manual zero adjustment ).
In
Programming mode
the key is used to increase a given value or switch to the next
menu or function.
Ð
When this key is pressed in
Measuring mode
the measured value on the numeric
display is reduced. A short pressure decrements by 1. If the key is held down the
measured value is rapidly reduced ( manual zero adjustment ).
In
Programming mode
the key is used to reduce a given value or to switch to the next
menu or function.
Ï
&
Ð
When
the
Ï
and
Ð
- keys are pressed simultaneously in
Measuring mode
either an
automatic zero adjustment or a gauge calibration is done.
If
the
Ï
and
Ð
keys are pressed simultaneously for more than 3 seconds the internal
zero adjustment offset value is erased and the mode for the mechanical adjustment of
inductive probes is started.
Refer to chapter 4.3 "Mechanical adjustment of inductive probes".
3.2 Foot switch functions
1. On manual gauge selection you switch to the next gauge
2. Output of measured value at RS232 output
3. Display control :
trAnS
→
transfer of measured value to display
hold
→
hold display value
4 Automatic zero adjustment or gauge calibration.
Notice :
The functions 2 ... 4 can be switched on and off in the Setup menu.