HyQuest Solutions Pty Ltd
Point Velocity Display - Model PVD100
©
Copyright
42
Issue 1.00 4 Apr, 2007
Appendix D Example Calibration Certificate
HYQUEST SOLUTIONS PTY. LTD., AUSTRALIA
------------------------------------------
CERTIFICATE OF CALIBRATION Date: 27/4/06
--------------------------
Current Meter Model: xxxxxx
Serial No. 06-999
Fan
No.
1
Serial
No.
06-123
Diameter:
125mm
Pitch:
0.25m
Type of Support:
Method of Calibration: Average Value Equation
n < 0.42 V = 0.2190 n +0.0153 m/s
0.42 < n < 3.73
V = 0.2459 n +0.0041 m/s
n > 3.73 V = 0.2508 n -0.0142 m/s
Starting Velocity = 0.025 m/s
Maximum Velocity = 5.000 m/s
Note: 'n' denotes the number of revolutions of the propeller per second
'V' is the water velocity in meters per second.
Date:
27/4/06
INSPECTED:__________________
_____________________________________________________________
| The above meter has been calibrated in accordance with
|
| AS 3778.6.3-1992. This standard is identical with and has |
| been reproduced from ISO 3455:2007.
|
| The clause below has been printed with the permission of |
| Standards Australia from AS 3778.6.1-1992. This standard is |
| identical with and has been reproduced from ISO 2537:2007. |
| Clause 7.6 Recalibration. Meters shall be recalibrated
|
| whenever their performance is suspect. In practice for
|
| individually rated meters, recalibration is sometimes
|
| carried out at yearly intervals or after 300 hours of use, |
| whichever is the shorter.
|
|_____________________________________________________________|
D127.08.00
Range Values
First Term
(Slope)
Second Term
(Intercept)