Programming for HIP Cleaning Solution Revision 2.02 Software
Before proceeding verify that the Hydrim L110W has revision 2.02 software. By pressing buttons P3
& P4 while powering the unit on the display shows “r” in the first digit followed by the first digit of
the revision number and then the last two digits. (For example if the revision is 2.02 then the display
will show r2 followed by 02). If the Hydrim L110W has 2.02 software proceed with the programming.
The Hydrim L110W software has been upgraded for use of the HIP (Hydrim Instrument Protection)
Cleaning Solution. Solution with enhanced instrument protection properties.
With 2.02 software, verify that the settings are correct for HIP Cleaning Solution. By pressing buttons
P1 & P3 while powering the unit on you will have access to the wash settings. By pressing P3 you can
select among the three wash settings c1, c2 and c3. In order to adjust the value of a setting, use the P0
button.
Dosing settings table
Wash setting “c2”
00
Prewash setting “c1”
01
Rinse setting “c3”
02
Note that there are values available other than those in the table. Please do not select these values
unless advised to do so by a SciCan service technician.
Important!
Ensure that all instruments are clean and free of stains and rust before using the HIP chemical.