15
Control Panel
Check software revision:
With unit OFF hold in buttons P3 & P4 turn the unit ON and release all
buttons. The display shows “r” in the first digit followed by the first digit of the revision number and
then the last two digits. (For ex. If the revision is 2.01 then it will display r2 followed by 01)
Water softener:
With the unit OFF hold in buttons P1 & P2, turn the unit ON and release all buttons.
The display shows the actual water softener setting value. Default value is 0. Adjust by pressing P0 to
decrease and P1 to increase. Turn the power OFF to save the selected setting.
Set drying time:
With the unit OFF hold in buttons P0 & P3, turn the unit ON and release all buttons.
The display shows “tl” followed by the wash time 5 minutes. Press P3 and the display shows “tE”
followed by the wash temperature 50
°
C. Press P3 and the display shows “dt” followed by the drying
time 10 minutes. Drying time can be adjusted from 10 to 20 minutes by pressing P0 to decrease and P1
to increase. Turn the power OFF to save the selected drying time.
Device test cycle:
With the unit OFF hold in buttons P2 & P3 turn the unit ON and release all buttons.
The unit starts filling the chamber then the dosing pump runs for 20 seconds.
Fault Codes
Fault #
Problem
Description
E1
Water heating failure
Water temperature < set point after timeout during
“Circulation and heating” phase. See E1 explanation.
E2
Filling failure
Timeout on filling up the chamber. See E2 explanation.
E3 Chamber
temperature
reading failure
Chamber temperature reading outside range
E4
Water evacuation failure
Timeout on water evacuation from chamber (60
seconds). See E4 explanation.
E6 Serial
transmission
failure
Control modules not communicating. Check wiring to
controllers. Replace Master Controller & Slave
Controller if needed.
E8 Water
inlet
temperature
reading failure
Water inlet temperature reading outside range. Check
water inlet temperature sensor (3.6K
Ω
).
E9
Program timeout
Timeout on finalizing the cycle. Check wash & drying
time. Call SciCan Technical Support
E0
Sidebag filling failure-
full pressure switch
blocked into ON position
Timeout on filling the heat exchanger (5 minutes).
Check pressure switch on Emitter Housing.
ED Dosing
system
failure Timeout on dosing. See ED explanation
EF
Flow error
Detergent flow switch defective. See EF explanation.
Chemical
Light
On steady cannot start
cycle
Check long green plug on Slave Controller.
Power Switch
Door Handle