MCS 1000 Series
Error analysis / remedy
HYDAC FILTER SYSTEMS GMBH
en(us)
Page 60 / 76
BeWa MCS1000 3435427f V130 en-us 2012-06-15.doc
2012-06-15
Error status 2 "active high"
The signal at the switching output is permanently free of voltage. The MCS has not
detected an error.
This error occurs also when external measurement voltage is switched off.
Only a superordinate controller can differentiate between this error signal and
normal operation without detection of a particle. Evaluate the signal duration and the
behavior of the two switching outputs S1 and S2 separately.
0 V
0
U
t
U
M1/2
Operation voltage
ON
Measurement voltage U
M1/2
OFF
Error status - "Device Ready"
In the "Device Ready" error status there is a signal change from "low" to "high"
0 V
0
U
M1/2
U
t