HDA 5500 for HLB 1000 and CS 1000
- 8 -
7.2 Overview of basic settings menu
Menu point
Setting
Setting range
Default setting
Device:
Device selected to display the values measured
by the HLB 1000 connected
Device selected to display the values measured
by the CS 1000 connected
HLB,
CS
HLB
When HLB 1000 is selected
Allocation switching output 1 (S.S.1)
Switching output 1 allocated to the signal
of rel. change in viscosity
Switching output 1 allocated to the signal
of rel. change in dielectric constant
Switching output 1 allocated to the signal
of
rel.
humidity
Switching output 1 allocated to the signal
of
temperature
r.ViS.
r.diE.
r.HuM.
TEMP.
r.ViS.
Switching function switching output 1 (S.M.1)
Switching output 1 set to switch point and
hysteresis mode
Switching output 1 set to switching window mode
SP
Win
SP
Switching direction switching output 1 (S.d.1)
Relay activated
Relay deactivated
on
oFF
on
Switch on delay switching output 1 (T.on1)
Time, in seconds, which must elapse once the particular switch
point has been reached or exceeded in order for a switching
function to occur.
0..99.99 s
0
Switch-off delay switching output 1 (T.oF1)
Time, in seconds, which must elapse once the value has fallen
below the particular switch-back point in order for a switching
function to occur.
0..99.99 s
0
Switching function switching output 2 (S.S.2)
Similar to switching output 1 (see above), and likewise for:
•
Switching mode (S.M.2)
•
Switching direction (S.d.2)
•
Switch-on delay (T.on.2)
•
Switch-off delay (T.oF.2)
r.ViS.
r.diE.
r.HuM.
TEMP.
SP
Win
on
oFF
0..99.99 s
0..99.99 s
r.ViS.
SP
on
0
0