HYDAC FILTER SYSTEMS GMBH
Operation | 7
BeWa FCU1210 3545830b en-us lq
57 / 92
Fig. 23:
Operating mode - Measurement at 15 … 345 bar
Make sure that the following sequence is observed:
ü
You have performed all of the steps listed in chapter
Putting the FCU into operation [
.
1.
Check the pressure at the measurement location. The
existing pressure must be in the range of 15 … 345 bar /
217 … 5000 psi.
WARNING! If the pressure is greater than 345 bar /
5000 psi, abort the measurement attempt.
2.
Screw the high-pressure adapter onto the INLET con-
nection of the FCU.