CTU in operation
Analysis chamber with basket mount
NOTICE
Ultrasonic tests without a suitable jig
The CTU will be damaged/destroyed
►
Do not reduce the fill quantity of the analysis liquid before a test.
►
Never put a test item into the analysis chamber without the
corresponding jig
►
Use the hook-in basket. Note the maximum weight of the test item of
0.5 kg.
The minimum fill quantity of the analysis liquid for ultrasonic testing is given
in the software by the defined minimum value of the fill quantity.
The temperature of the ultrasonic bath is continuously monitored. If a
temperature of 40°C is reached during ultrasonic testing, the monitor will
switch off the ultrasonic vibrator. Testing will be interrupted while the unit
cools down.
Close the ball valve on the filter membrane holder during the ultrasonic test.