CTU in operation
ConTeS Software
The main window of the ConTeS appears automatically after the PC has
been switched on and has started up.
For details on the ConTeS software, please see the
"ConTeS operating
instructions".
Clean-Room Conditions in the Analysis Chamber - Rinsing of the inner
chamber
Clean the analysis chamber of the CTU before the beginning of a
contamination test series with the "Rinsing of inner chamber" program
defined by the volume or by the time/duration.
During operation, as well as between extractions, make sure that the analysis
chamber is always closed and is opened for only a short time for the insertion
of the component to be analyzed. Only in this way can you ensure that
contamination from the surrounding area is as minimal as possible.