Adjusting the counter balance valve
3. Screw the pressure gauge (B)
onto the gauge port.
4. Switch the unit on.
Check the suction intake
behavior of the pump and the oil
flow.
If, from the noise of the pump or
the outlet side (OUT), you can
hear/identify an even flow rate,
please proceed.
5. The pressure gauge (B) at the
gauge port shows the pressure
at the input of the
ContaminationSensor
(max. 60 bar)
6. Unscrew the cap nut from the
counter balance valve
.
7. Release the lock nut on the
counter balance valve.
CSM-E
en(us)
Page 36 / 64
MoWa CSM-E 3911629a en-us 2017-12-11.docx
2017-12-11
Summary of Contents for CSM-E
Page 54: ...Technical Data CSM E en us Page 54 64 MoWa CSM E 3911629a en us 2017 12 11 docx 2017 12 11 ...
Page 62: ......
Page 63: ......