User Manual
Rev 9.0
Page 35
HV Diagnostics
Step
Procedure
(Run Automatic Test )
Example
AR 1:
Use Last Seq. or
Start New Test
To repeat the previous test sequence:
•
Select “Use Last Test Sequence” from “Main
Menu”
•
Proceed to Step AR 5
Otherwise, select:
•
“Start a new Test”
AR 2:
De / activate
Reporting
To activate reporting:
•
Select “YES”, See 6 Reporting
To conduct a test without generating a report:
•
Select “NO”
AR 3:
(If reporting
active: this step
follows report
setup
completion)
To run an test sequence:
•
Select “Auto Seq”
AR 4:
Select Sequence
•
Select the appropriate test sequence
•
To continue, select “NEXT”
AR 5:
Parameter
Verification
•
Verify that the selected sequence defines the
correct test parameters
To run the auto test sequence:
•
Select “START” from the bottom of the screen
AR 6:
HV Activation
Once the activation screen appears,
•
Press the HV switch (6) within 10 seconds.
If the HV switch is not activated within the 10
second window, the “Start Auto Test Seq. “screen
will reappear.
AR 7:
Test Start up
“Startup” appears on the screen to indicate that the
HVA is initializing test
USB
Use Last Test Seq.?
3.0kV Maint. C. HYBR (T)
(T)
Start a New Test
Manual Mode
Reports & Setup
13.3.2009 10:23
Main Menu
Start a New Test
Do you want to
store a report?
Yes
No
MAIN
Start a New Test
Use an existing
Auto Test Sequence
or use Manual Mode?
Auto Seq.
Manual
MAIN PREV
Select Auto Test Seq.
3.0kV Maint. C. HYBR (T)
1.0kV Maint. C. XLPE (T)
BACK SELECT NEXT
Start Auto Test Seq.
1.0 kV Cable XPLE
Maintenance Test Sequence
Test Voltage: 1.0kVrms
Duration: 15 min
Waveform: Sine 0.1Hz
Trip out on
Reporting Active
MAIN BACK
START
ATTENTION
High Voltage!
Press I/O Button to
switch ON High Voltage
Auto Test Sine 0.1Hz
STOP
T: 00:00 / 15min
1.0
kV
rms
0.0kV
0.0µA
Startup