Switches and Controls
Pos. Name
Description
1
Emergency OFF
Activates emergency shutdown.
Operation is only possible when Emergency OFF is deactivated.
•
To Activate Emergency Off
Press in
•
To Deactivate Emergency Off
Rotate and Release latch
2
LED green
Indicates HV status.
Green light indicates
High Voltage is OFF
3
LED red
Indicates HV status.
Red light indicates
High Voltage is ON (possible DANGER)
DUT not discharged
(residual voltage >100V)
4
Display screen
Displays menu, options and status information.
5
Navigation knob
Enables user to select options and functions shown on display.
See 4.2 User Interface
•
To scroll selection up or down
Rotate
•
To enter selection
Click (push in)
6
HV switch
[on/off]
Activates high voltage.
•
To activate HV output
Press within 10 seconds after START
See 5.2 Manual Test Procedure: Step MR 2;
See 5.3 Automatic Test Procedure: Step AR 6
7
Key switch
[on/off]
Locks the unit to prevent against unauthorized use.
•
To disable unit
Remove key from the OFF Position
•
To reactivate unit
Replace key and turn to ON Position.
Front Panel
1
2
3
4
5
6
7