User Manual
HV Diagnostics
Page 30
Rev 9.0
6.3
Automatic Test Mode
This HVA test mode allows automatic testing of a DUT according to a
specific sequence in order to satisfy specific requirements (e.g. IEEE,
IEC standards). The test sequence can be configured, modified and
saved at any time before testing.
Configuring Auto Test. Sequence- Detailed Steps
NOTICE
No Automatic Test Sequence programmable for Sheath/Jacket Test and
Sheath Fault Location.
Already programmed in Manual Mode!
Main Menu
USB
Use Last Test Seq.?
3.0kV Maint. C. HYBR (T)
(T)
Start a New Test
Manual Mode
Reports & Setup
13.3.2009 10:23
Reports &Setup
Reports & Setup
View Reports (50)
Store Rep. to USB
Edit Auto Test Seq
.
Instrument Settings
Display Contrast 4
Service
MAIN
Edit or New?
Edit Auto Test Seq.
3.0kV Maint. C. HYBR (T)
BACK SELECT
EDIT DEL NEW
Edit Auto Test Seq.
3.0kV Maint. C. HYBR (T)
BACK SELECT EDIT DEL
NEW
Set Sequence
Save Sequence
Edit Auto Test Seq.
3.0 kV Cable HYBR
Maintenance / Sine 0.1Hz
Test Voltage: 1.0kVrms
Duration 15min
Trip out on Arc
Finished
Save or Cancel
CANCEL PREV
SAVE
New Auto Test Seq.
1.0 kV Cable XPLE
Maintenance / Sine 0.1Hz
Test Voltage: 1.0kVrms
Duration 15min
Trip out on A
rc
Finished
Save or Cancel
CANCEL PREV
SAVE
HVA guides user through series of questions
•
DUT Type
•
Voltage Rating
•
(Cable) Insulation
•
Test Purpose
•
Waveform
•
Frequency
•
Voltage Steps
•
Test Voltage (for each step)
•
Duration (for each step)
•
Arc Management Mode
Main Menu