![Huvitz HDR-9000 Service Manual Download Page 61](http://html1.mh-extra.com/html/huvitz/hdr-9000/hdr-9000_service-manual_2190280061.webp)
- 60 -
‘Ref.Body’, ‘Motor No’, ‘Reference’ and ‘Sensor’.
(Dial) : changes option value.
: start executing auto-detection of motor origin.
The “Ref.Body” option gives a chance to select the target refractor, right or left.
The “Motor No” option is given to select the target motor in the selected refractor including all the
motors for driving 6 different disks, 3 different axis, pd and convergence.
The “Reference” item shows the amount of pulse count input by the operator to the selected target
motor.
The “Sensor” item shows the read value from the sensor detecting the origin of selected target
motor. The information in the parenthesis indicates recognition status and recognition strength of
sensor. “ON” means that the origin of target motor is detected. “OFF” means that the origin of
target motor is not detected yet. For example, “OFF-24” means that the origin is not detected and
it’s off from the origin as much as 24 of sensor value.
Summary of Contents for HDR-9000
Page 1: ...Service Manual Digital Refractor HDR 9000 HUVITZ Co Ltd...
Page 2: ...1...
Page 9: ...8 Engineer Mode Refer to appendix for more detailed information about this mode...
Page 17: ...16 Unscrew M2 6 screws and remove BACK CAP BACK CAP...
Page 18: ...17 Unscrew 3 M2 6 screws and remove TOP FRONT CASE TOP FRONT CASE...
Page 19: ...18 Unscrew 6 M2 6 screws and remove INNER CASE L INNER CASE R INNER CASE L INNER CASE R...
Page 30: ...29 4 6 Removing the LCD Panel Cover of the Operation Panel...
Page 32: ...31 4 8 Disassembling the Junction Box...
Page 37: ...36 5 1 2 Refractor Body Wiring Diagram...
Page 40: ...39 5 1 3 Junction Box Wiring Diagram...
Page 42: ...41...