Replacing HALL sensor
-
Before detaching a PI sensor, mark the position of the sensor along its boundary with
a sharp tool such as a cutter knife. Then you can assemble a new sensor at the most
close position with the original one.
-
Disconnect the connector of the HALL sensor board.
-
Unscrew and remove the HALL sensor board.
-
Screw a new HALL sensor board slightly at the position where a defective board’s
been removed.
-
By referring the marks of original position of replaced sensor board, fix the new
board firmly.
-
Connect the connector of the new sensor board.
-
Turn on the power and check if the initialization sequence is performed correctly. All
the open apertures of the disks should be aligned concentric after the initialization
sequence. Otherwise contact the customer service department at Huvitz for further
measurement and instructions.
-
Execute the Engineer Menu mode on the Operation Panel.
-
Execute the Test Mode function.
-
Check the PI sensor value and if the sensor is detecting its origin correctly.
<Adjusting the position of HALL sensor>
Summary of Contents for HDR-9000
Page 1: ...Service Manual Digital Refractor HDR 9000 HUVITZ Co Ltd...
Page 2: ...1...
Page 9: ...8 Engineer Mode Refer to appendix for more detailed information about this mode...
Page 17: ...16 Unscrew M2 6 screws and remove BACK CAP BACK CAP...
Page 18: ...17 Unscrew 3 M2 6 screws and remove TOP FRONT CASE TOP FRONT CASE...
Page 19: ...18 Unscrew 6 M2 6 screws and remove INNER CASE L INNER CASE R INNER CASE L INNER CASE R...
Page 30: ...29 4 6 Removing the LCD Panel Cover of the Operation Panel...
Page 32: ...31 4 8 Disassembling the Junction Box...
Page 37: ...36 5 1 2 Refractor Body Wiring Diagram...
Page 40: ...39 5 1 3 Junction Box Wiring Diagram...
Page 42: ...41...