Procedure
To begin the aging test, press the following key sequences on the Operation Panel.
ALT + MENU
ALT + START
ALT + NEXT
3.2.2 Post Aging Test Procedures for the Refractor Body
①
Check the PD adjustment function.
Press PD button.
Check if the PD-check lenses are inserted properly.
Turn the dial clockwise or counter-clockwise several times.
Press PD button again to finish this step.
②
Check the convergence function.
Press FAR and NEAR button on the display several times.
Check if the Refractor Body is converging correctly.
Stop this step when the HDR-9000 Refractor Body is in FAR mode.
③
Check the initialization sequences again.
Turn off the power.
Turn on the power again.
Check if the disk plates in the Refractor Body are aligned properly after the initialization
sequences.
Repeat this step 3 to 5 times.
④
Check the auxiliary lenses.
To enter the lens-checking mode, press the following key sequences.
ALT + MENU
ALT + START
ALT + TEST
"Check Lens"
Change the selection of eye to binocular:
By pressing the OU button.
Summary of Contents for HDR-9000
Page 1: ...Service Manual Digital Refractor HDR 9000 HUVITZ Co Ltd...
Page 2: ...1...
Page 9: ...8 Engineer Mode Refer to appendix for more detailed information about this mode...
Page 17: ...16 Unscrew M2 6 screws and remove BACK CAP BACK CAP...
Page 18: ...17 Unscrew 3 M2 6 screws and remove TOP FRONT CASE TOP FRONT CASE...
Page 19: ...18 Unscrew 6 M2 6 screws and remove INNER CASE L INNER CASE R INNER CASE L INNER CASE R...
Page 30: ...29 4 6 Removing the LCD Panel Cover of the Operation Panel...
Page 32: ...31 4 8 Disassembling the Junction Box...
Page 37: ...36 5 1 2 Refractor Body Wiring Diagram...
Page 40: ...39 5 1 3 Junction Box Wiring Diagram...
Page 42: ...41...